Hit the right concentration


If the concentration in your sample is too high, we dilute it. If the concentration is too low, we concentrate your sample. In either case, we try to hit the target of 0.5–4 million cells/mL. See the table below for our recommendations:
5 · 107 cells/mL
10-2
Preferred Dilute 1:100, alternatively Dilute 1:1 000 if measurement gives a conductivity error.
5 · 106 cells/mL
1:10
Dilute 1:100. Contact SBT for guidance or review Hit the right conductivity if you receive conductivity errors.
≤5 · 104 cells/mL
N/A
Generally it is not recommended to use BactoBox for very dilute samples. For workarounds, see Concentrate your sample.
You can do combinations to get even high dilution factors. Just like how we do three Dilute 1:100 steps in a row to get a 1:1 000 000 dilution overall.
Quick 1-step dilutions
The above dilution principles rely on integers divisible by 10. This straightforward method typically works well for samples with high bacterial concentrations. However, there are situations where these dilutions may not be optimal.
For highly conductive broths, some cultivation media may require a dilution greater than 1:100 to achieve acceptable conductivity levels. In such cases, performing a 1:201 dilution by adding 50 µL of the sample to 10 mL of diluent might be beneficial.
For rapid and rough qualitative assessments, a 1-step dilution is efficient and conserves consumables. For example, you can mix 10 µL of your sample with 10 mL of diluent to achieve a 1:1 001 dilution.
Ensure precise pipetting to maintain accuracy, especially when handling small volumes.
80 µL
10 mL
126
50 µL
10 mL
201
40 µL
10 mL
251
25 µL
10 mL
501
10 µL
10 mL
1 001
8 µL
10 mL
1 251
Very low bacterial concentration?
BactoBox® is generally not suitable for very low bacterial concentrations.
Motivation
BactoBox® counts individual particles. If there are too few particles, the concentration estimate is too uncertain. If there are too many particles, we can't tell them apart. Therefore, we have limit of detection (LoD). BactoBox® outputs concentrations in these ranges:
30 000–5 000 000 cells/mL (3·10⁴ to 5·106 cells/mL)
30 000–5 000 000 total/mL (3·10⁴ to 5·106 total/mL)
See limits of detection for more details.
Last updated