ruler-combinedTechnical specification

Parameter
Value

Dimensions

240×140×60 mm

Net weight

1 200 g

Input voltage

12 V⎓ / 2 500 mA / 30 W

Environment temperature for operation

15–35 °C

circle-info

Normal BactoBox® operating temperatures is 20 - 25 °C. Broader temperature range is only possible if a custom diluent is used. If you are operating BactoBox® in an anaerobic chamber, the temperature is often elevated. This requires a reduction in PBS. Target a diluent conductivity of 1850 µS/cm at the desired temperature. Use the built-in conductivity program to test the conductivity.

temperature-quarter For lower temperatures, increase the proportion of PBS.

temperature-full For higher temperatures, decrease the proportion of PBS.

Power usage

BactoBox® power usage in various situations.

Situation
Average
Peak

Idle

3.60 W

3.84 W

Measure

6.00 W

13.9 W

circle-info

The power usage peaks whenever the flow changes direction or quickly accelerates. These peaks usually last for a very short duration.

Last updated