Consumables
List of consumables for BactoBox®. See also Vials, flasks, and liquids.
Flow cell
Contains the microfluidic flow paths where we measure bacteria.
See Install flow cell for installation instructions.
Clean the flow cell before you remove it
It is good practice to run the Clean program with a disinfection vial before you remove the flow cell.
Replace the external filter when you replace the flow cell
It is good practice to replace the external filter whenever you replace the flow cell. See Install tubing kit for details.
Qualify your BactoBox® setup when you replace the external filter.
The flow cell is worn out after 250 measurements
After 250 measurements, the flow cell reaches its usage limit and is automatically locked. This limit is set to guarantee the factory calibration of the flow cell.

Item numbers
FC03
Contact SBT to purchase This bundle includes a flow cell and external filters. Remember to change the external filter when you change the flow cell.
Not possible. The flow cell is patented.
External filter
Placed between the BactoBox® device and the tubing kit.
Pore size is 40 µm.
Replace the external filter when you replace the flow cell
It is good practice to replace the external filter whenever you replace the flow cell. See Install tubing kit for details.
Qualify your BactoBox® setup when you replace the external filter.


PF01
) that contains multiple external filters.Item numbers
Not recommended.
Syringe filter
Use together with the dilution dispenser (see Accessories).
Replace the syringe filter at the start of each day
Even within a single day, the syringe filter clogs after repeated use. If it feels like you need to press harder to dispense, replace the filter. Most filters clog after about 50 dispenses.

Item numbers
Not recommended.
Cell sieve (10 µm)

Item numbers
Not recommended.
Glass beads 3 mm (50 g)
See Bead beating.

Item numbers
Not recommended.