Flow cell
Use with BactoBox® to measure bacteria

🏷️ Individual item number 20030
🏷️ Bundled as FC03
⛽ Lasts 250 measurements
Expires 2 years from production date
Once opened, use before 6 months
Replace the external filter when you replace the flow cell.
Not sold individually
Purchase FC03 that comes with 2 × external filters.
FC03
This bundle includes a flow cell and two external filters. Remember to change the external filter when you change the flow cell.
Third party
Not possible The flow cell is patented by SBT.
Use cases
Install the flow cell in your BactoBox®.
Replace the flow cell in these cases:
After 250 measurements. The BactoBox® software automatically prompts you to do this.
Clean the flow cell before you remove it.
Replace the external filter when you replace the flow cell
It is good practice to replace the external filter whenever you replace the flow cell. See Install tubing kit for details.
Run QC test when you replace the external filter.
The flow cell is worn out after 250 measurements
After 250 measurements, the flow cell reaches its usage limit and is automatically locked. This limit is set to guarantee the factory calibration of the flow cell.