Use with BactoBox® to measure bacteria
envelopeContact SBT to purchase
🏷️ Individual item number 20030
20030
🏷️ Bundled as FC03
FC03
⛽ Lasts 250 measurements
Expires 2 years from production date
Once opened, use before 6 months
Replace the external filter when you replace the flow cell.
Purchase FC03 that comes with 2 × external filters.
This bundle includes a flow cell and two external filters. Remember to change the external filter when you change the flow cell.
Third party
Not possible The flow cell is patented by SBT.
M
Flow cell bottom view with electrode pads, memory chip, and flow channels for the liquid
N
Flow cell top view where you can see the flow cell ID
Install the flow cell in your BactoBox®.
Replace the flow cell in these cases:
After 250 measurements. The BactoBox® software automatically prompts you to do this.
Clean the flow cell before you remove it.
It is good practice to replace the external filter whenever you replace the flow cell. See Install tubing kitarrow-up-right for details.
Run QC test when you replace the external filter.
After 250 measurements, the flow cell reaches its usage limit and is automatically locked. This limit is set to guarantee the factory calibration of the flow cell.