Deep clean vial
Use to deep clean your BactoBox® setup.

DC02🏷️ Individual item number 50055
🏷️ Bundled as DC02
💧 Contains 12 mL
🌡️ Store at 15–25℃
Expires 2 years from production date
Once opened, use before end of day
The deep clean vial contains 1 % pepsin in hydrochloric acid (0.1 % HCL).
DC02
Not recommended DIY deep clean vials may irreparably damage your BactoBox® or flow cell.
Use cases
Use to deep clean your BactoBox®.
Deep clean your BactoBox® in these cases:
At least once a week. Use the 15-minute option.
To recover a flow cell that failed the QC test. Try with the 15-minute option first. If the issue persists, try the 10-hour option.
See also deep clean explained.
Gallery

What is a vial?
In SBT terms, a vial is a 15 mL conical centrifuge tube.
BactoBox® connects to a vial via the tubing kit. See the vial rack for a convenient way to store vials.
GHS

H334 May cause allergy or asthma symptoms or breathing difficulties if inhaled.
Last updated
Was this helpful?