> For the complete documentation index, see [llms.txt](https://help.sbtinstruments.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://help.sbtinstruments.com/everyday-use/run-a-sample.md).

# Run a sample

Disperse the cells, dilute into the working range, measure, and clean.

Your BactoBox® counts accurately between 30 000 and 5 000 000 cells/mL (3 × 10^4 to 5 × 10^6) in the vial you measure. Read [Before you measure](/everyday-use/before-you-measure.md) first.

## Step by step

{% stepper %}
{% step %}

### Disperse cells

Cells in a culture tend to stick to each other. To get repeatable dilutions, disperse the cells before your first dilution.

Transfer some culture to a tube you can vortex, such as a microcentrifuge tube or a centrifuge tube. Vortex for 30 seconds at maximum speed.

Some organisms form chains or clumps that a vortex cannot separate. If yours does, see [Break up cell clumps](/more/how-to-guides/break-up-cell-clumps.md).
{% endstep %}

{% step %}

### Dilute 1:100

Transfer 101 µL from the tube into a diluent vial and [rinse the tip](/more/how-to-guides/basic-sample-preparation.md) in the vial. Vortex the vial for 10 seconds at maximum speed.

A 1:100 dilution suits young cultures, from lag to early exponential phase.

<figure><img src="https://4216107837-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FeYTtd7c0A325muFr5lMW%2Fuploads%2FMRVuIftr25wf36WFavtS%2Fillustration-1%3B100-new.png?alt=media&#x26;token=8ef3758f-5884-487c-b009-b5cd8e13edd3" alt="" width="519"><figcaption></figcaption></figure>
{% endstep %}

{% step %}

### Dilute again to 1:10 000, for dense cultures

Overnight and stationary cultures are too dense for 1:100. Transfer 101 µL from your 1:100 vial into the second diluent vial and [rinse the tip](/more/how-to-guides/basic-sample-preparation.md) in the vial. Vortex the vial for 10 seconds at maximum speed.

Culture in mid-exponential phase? Use 1:1 000 instead. See [Hit the right concentration](/more/how-to-guides/hit-the-right-concentration.md).

<figure><img src="https://4216107837-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FeYTtd7c0A325muFr5lMW%2Fuploads%2FRGiAG2OpasfkOGwzot2u%2Fillustration-1%3B10%2C000-new.png?alt=media&#x26;token=bffbfa0f-d367-472a-ab8d-de6f0f9aef21" alt="" width="519"><figcaption></figcaption></figure>
{% endstep %}

{% step %}

### Measure

Move the tubing kit to your final vial and press <kbd>Measure</kbd>. Multiply the result by the dilution factor.

<figure><img src="https://4216107837-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FeYTtd7c0A325muFr5lMW%2Fuploads%2FLFGAxYPqlj6NQ7qq5XC3%2Fbloodstone-direct-measure.avif?alt=media&#x26;token=5cd7797e-f60c-4f93-8a6b-55fb5ce74920" alt=""><figcaption></figcaption></figure>

{% hint style="info" %}
Result outside the working range? See [Hit the right concentration](/more/how-to-guides/hit-the-right-concentration.md).
{% endhint %}
{% endstep %}

{% step %}

### Clean

Move the tubing kit to your disinfection vial and press <kbd>Clean</kbd>. Do this after every measurement.

<figure><img src="https://4216107837-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FeYTtd7c0A325muFr5lMW%2Fuploads%2FT8k2R8qYsX9dd9Z73SI1%2Fbloodstone-direct-clean.avif?alt=media&#x26;token=ff100f0b-49d1-47cc-9a18-b5b23a455e47" alt=""><figcaption></figcaption></figure>
{% endstep %}
{% endstepper %}


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://help.sbtinstruments.com/everyday-use/run-a-sample.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.