You need these items:
1 × BactoBox® setup
1 × deep clean vial
Transfer the tubing kit to the deep clean vial.
Press and hold Measure for 2 seconds to enter the Programs menu.
Use the navigation buttons (OK ← ∨ ∧) to select the Deep clean program.
Press OK to continue.
Use the navigation buttons (OK ← ∨ ∧) to select the Deep clean duration.
Press OK to start the Deep clean program.
The Deep clean program is done. Discard the deep clean vial. It is for single use only.
Perform a QC test after the Deep clean program is done.
Deep clean your BactoBox® in these cases:
At least once a week. Use the 15-minute option.
To recover a flow cell that failed the QC test. Try with the 15-minute option first. If the issue persists, try the 10-hour option.
See Maintenance for more information.
See Deep clean explained for additional detail.