Prepare for first use

Deep clean, then run a QC test. About 25 minutes, before the first sample.

Getting started: Unboxing and Installation are done. This page is Prepare for first use. Run a sample comes next.

First use always starts with a deep clean, followed by a QC test. The QC test also unlocks the new flow cell for measurements.

Deep clean with a deep clean vial for 15 minutes. Then the QC test: clean with a disinfection vial, measure the QC vial, clean with the same disinfection vial, and measure a dilution vial as the blank. Two check marks mean BactoBox is ready for samples.

Get things ready

All four come in the Getting started bundle.

Deep clean for 15 minutes

Place the deep clean vial in the vial rack and move the tubing kit to it.

Three steps. Unscrew the cap of the deep clean vial and keep it for later. Place the open vial in one of the holes of the vial rack. Move the tubing kit to the vial. Its plug replaces the cap, and the steel tubes reach down into the liquid.
  1. Press and hold Measure for 2 seconds, until the Programs menu appears.

  2. Use the navigation buttons (∨ ∧) to select Deep clean, and press OK.

  3. Select 15 minutes and press OK to start.

Selecting 15 minutes in the Deep clean program

When the program is done, close the deep clean vial with its screw cap. Discard it as biological waste.

Run the QC test

1

Get a QC vial

Cut out one ready-made QC vial (QC07). Cut out only one at a time.

Cutting one QC vial off the strip

Using QC stock (QC04, QC05 or QC06)? Make a QC vial from QC stock first.

2

Start the QC test program

  1. Press and hold Measure for 2 seconds, until the Programs menu appears.

  2. Press Measure again briefly to select QC test. In this program, Measure works as OK.

  3. Set the target concentration with the buttons. On a ready-made QC vial, it is printed in the center of the sticker. For QC stock, it is on the certificate of analysis.

Setting the target concentration in the QC test program
3

Clean

Move the tubing kit to the disinfection vial and press Measure (OK). If you just ran a Clean, BactoBox® skips this step.

4

Measure the QC vial

Shake the QC vial vigorously for 10 seconds, because the QC beads settle over time. Move the tubing kit to it and press Measure (OK). Press Measure (OK) again to leave the result screen.

Measuring the QC vial
5

Clean again

Move the tubing kit back to the same disinfection vial and press Measure (OK). This rinses out the QC beads.

Cleaning with the disinfection vial
6

Measure the blank

Move the tubing kit to the ready-made dilution vial, with nothing added, and press Measure (OK).

Measuring the dilution vial as the blank
7

Ready for samples

Two check marks on the screen mean the QC test passed. BactoBox® is ready for samples. The QC test also calibrates CIZE for the mounted flow cell.

Discard the QC vial and the dilution vial. Both are for single use. Keep the disinfection vial for the Clean after each sample.

Did the QC test fail?