Prepare for first use
Deep clean, then run a QC test. About 25 minutes, before the first sample.
First use always starts with a deep clean, followed by a QC test. The QC test also unlocks the new flow cell for measurements.
Get things ready
All four come in the Getting started bundle.
1 × deep clean vial
1 × QC vial
1 × dilution vial
Deep clean for 15 minutes
Place the deep clean vial in the vial rack and move the tubing kit to it.
Press and hold Measure for 2 seconds, until the Programs menu appears.
Use the navigation buttons (∨ ∧) to select Deep clean, and press OK.
Select 15 minutes and press OK to start.

When the program is done, close the deep clean vial with its screw cap. Discard it as biological waste.
Run the QC test
Get a QC vial
Cut out one ready-made QC vial (QC07). Cut out only one at a time.

Using QC stock (QC04, QC05 or QC06)? Make a QC vial from QC stock first.
Start the QC test program
Press and hold Measure for 2 seconds, until the Programs menu appears.
Press Measure again briefly to select QC test. In this program, Measure works as OK.
Set the target concentration with the buttons. On a ready-made QC vial, it is printed in the center of the sticker. For QC stock, it is on the certificate of analysis.

Clean
Move the tubing kit to the disinfection vial and press Measure (OK). If you just ran a Clean, BactoBox® skips this step.
Measure the QC vial
Shake the QC vial vigorously for 10 seconds, because the QC beads settle over time. Move the tubing kit to it and press Measure (OK). Press Measure (OK) again to leave the result screen.

Clean again
Move the tubing kit back to the same disinfection vial and press Measure (OK). This rinses out the QC beads.

Measure the blank
Move the tubing kit to the ready-made dilution vial, with nothing added, and press Measure (OK).

Ready for samples
Two check marks on the screen mean the QC test passed. BactoBox® is ready for samples. The QC test also calibrates CIZE for the mounted flow cell.
Discard the QC vial and the dilution vial. Both are for single use. Keep the disinfection vial for the Clean after each sample.
Next: Before you measure.