TutorialNo skill requirement

Run QC test

Get things ready

You need these items:

  • 1 × disinfection vial

  • 1 × QC vial (see Make QC vial)

  • 1 × certificate of analysis for the QC vial

  • 1 × dilution vial (see the info box below)

Difference between QC test and qualification

When you qualify your BactoBox® setup, always use a ready-made SBT dilution vial.

For the regular weekly QC test, use a dilution vial from the dilution route you use for your own samples:

Besides the choice of dilution vial, qualification and QC test are the same.

Step by step

1

Start the QC test program

You do this with three button presses.

Enter the Programs menu

First, press and hold Measure until the Programs menu appears.

Press and hold Measure until the Programs menu appears.

Enter the QC test screen

Second, press Measure again briefly to select the QC test program.

Press Measure again briefly to select QC test program.
2

Set the target concentration

Check the sticker on the QC stock or the certificate of analysis and find the value for Concentration in prepared 1x QC test sample.

In this example, the value is 2 080 000 particles/mL (highlighted with a green circle).

Input the target concentration in million particles/mL.

Enter the value in million particles/mL

Press OK to submit the target concentration.

Does your QC test program look different?

Update your BactoBox® software. We introduced the Input target concentration screen in software v2025.10.

3

Clean the setup

Transfer the tubing kit to the disinfection vial.

Press OK to start the Clean program.

If you already followed Clean, BactoBox® automatically skips this step.

4

Measure on QC vial

Transfer the tubing kit to the QC vial.

Press OK to start the measurement.

Status: Pass

Status: Fail

5

Clean the setup

Transfer the tubing kit to the disinfection vial and press OK to clean.

6

Measure on dilution vial

Use the dilution vial with nothing added to it. This blank checks that neither your setup nor your dilution liquid adds a background count.

Transfer the tubing kit to the dilution vial and press OK to measure.

Status: Pass

Status: Fail

7

That is it!

Discard the QC vial and the dilution vial. They are for single use only.

When to perform a QC test

Perform a QC test in these cases:

  • At least once a week.

  • After a deep clean (of any length).

  • When you change your BactoBox® setup. Example: You replace the flow cell, the external filter or the tubing kit.

  • As part of qualification.

See Maintenance for more information.