Verify that your BactoBox® setup works as intended
This page teaches you how to qualify your BactoBox® setup. Qualification verifies that your setup works as intended.
Once you are done, you have a qualified BactoBox® setup. You can now measure your own samples.
Run a QC test. Use only SBT dilution vials (bundled as DV01).
DV01
This is different from a regular (weekly) QC test where the choice of dilution vial depends on your use case.
SBT dilution vials maintain a high quality standard, as these vials undergo rigorous quality checks.
See the QC test tutorial for next steps:
Qualify your BactoBox® in these cases:
At least once a year.
When you relocate your BactoBox® setup. Example: You install your BactoBox® for the very first time. Example: You ship your BactoBox® to another physical location. Example: You get your BactoBox® back from service/repair at SBT.
See Maintenance for more information.