> For the complete documentation index, see [llms.txt](https://help.sbtinstruments.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://help.sbtinstruments.com/inoc/workflows/get-an-active-culture-the-next-day/from-cryo-stock.md).

# From cryo stock

Find the volume of cryo stock that gives an active culture the next morning.

Cryo vials from the same batch hold about the same number of cells. So the same volume of stock gives the same seed each time. Find the right volume once, and use it from then on.

## Step by step

{% stepper %}
{% step %}

### Prepare 3 or 4 flasks in the afternoon

Use the medium, flask and volume you always use for your overnight culture.

<figure><img src="https://3226621149-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FuOGB3JuT2RALmbnviT9H%2Fuploads%2FxiddUOvfVmAUHgW9Q3Bi%2Fprepare-flasks.svg?alt=media" alt="Four flasks with medium in the afternoon: your usual medium, flask and volume."><figcaption></figcaption></figure>
{% endstep %}

{% step %}

### Make a dilution series of the stock

Thaw a cryo vial. Make 1:100 dilutions in growth medium, 10 µL into 990 µL each time. Vortex each tube before you take from it.

<figure><img src="https://3226621149-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FuOGB3JuT2RALmbnviT9H%2Fuploads%2FZFAz93q6kTYwMHWFfAFI%2Fcryo-dilution-series.svg?alt=media" alt="A thawed cryo vial and three tubes. 10 µL goes from each into the next, giving 1:100, 1:10 000 and 1:1 000 000. Each tube holds 990 µL medium."><figcaption></figcaption></figure>
{% endstep %}

{% step %}

### Seed each flask from a different tube

Put the same volume, for example 100 µL, into every flask. Take it from the stock for flask 1, and from the next tube for each next flask.

<figure><img src="https://3226621149-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FuOGB3JuT2RALmbnviT9H%2Fuploads%2FS768kEH0vweNnKNswuX4%2Fcryo-seed-flasks-v2.svg?alt=media" alt="100 µL from the stock and from each tube goes into its own flask. Each flask gets 100 times less stock than the one before."><figcaption></figcaption></figure>

Each flask gets 100 times less stock than the one before.
{% endstep %}

{% step %}

### Incubate overnight

Incubate as you always do. Slower bacteria may need longer. Use the incubation time that fits when you need the culture.

<figure><img src="https://3226621149-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FuOGB3JuT2RALmbnviT9H%2Fuploads%2FacT0Cq7VAEJs7q9XW53V%2Fincubate-overnight.svg?alt=media" alt="Flasks go into the incubator in the afternoon. In the morning, heavier seeds are more turbid."><figcaption></figcaption></figure>
{% endstep %}

{% step %}

### Measure every flask in the morning

Measure each flask with BactoBox® at the time you would use the culture.

<figure><img src="https://3226621149-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FuOGB3JuT2RALmbnviT9H%2Fuploads%2F99EdLCJjCpPplGmQvRLJ%2Fmeasure-flasks-v3.svg?alt=media" alt="Vortex each flask. Dilute each into its own dilution vial, the denser flasks further, from 1:10 000 to 1:100. Measure them one by one on BactoBox. Access lists cells/mL and CIZE for each flask."><figcaption></figcaption></figure>

See [Hit the right concentration](https://help.sbtinstruments.com/more/how-to-guides/hit-the-right-concentration.md) for the dilution.
{% endstep %}

{% step %}

### Pick the phase you want

Together, your flasks show the growth curve at the time you use the culture. The heavier seeds end at the same cell count, because they ran out of nutrients. The lighter seeds are still growing. CIZE tells the phases apart. Cells on the plateau are small, and growing cells are larger.

Our *E. coli* test shows what this looks like.

<figure><img src="https://3226621149-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FuOGB3JuT2RALmbnviT9H%2Fuploads%2FFRE8jVF7bpCvUhwom4gv%2Fpick-the-phase.svg?alt=media" alt="Five flasks measured the same morning, plotted on one growth curve. The two lightest seeds are still growing, the middle one is slowing, and the two heaviest are on the plateau. CIZE falls from 1.22 to 0.87 µm from the lightest to the heaviest seed."><figcaption><p>Cell counts and CIZE after about 11 hours in an <em>E. coli</em> test from cryo stock, with seeds 10 times apart.</p></figcaption></figure>

See [Know the growth phase of your culture](https://help.sbtinstruments.com/inoc/explainers/know-the-growth-phase-of-your-culture.md).

All flasks at the same count? They all reached the plateau. Seed lower next time.
{% endstep %}

{% step %}

### Use your seed from now on

Seed your overnight culture with that volume of stock. In the morning, measure it with BactoBox®, then dilute to your target and dose.

<figure><img src="https://3226621149-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FuOGB3JuT2RALmbnviT9H%2Fuploads%2FKvtKydNZIbxQ07hOOsPv%2Fcryo-routine.svg?alt=media" alt="Every afternoon, seed your volume of stock. Incubate overnight. Every morning, measure, dilute to target and dose. Test again when the stock batch, medium, flask or incubation time changes."><figcaption></figcaption></figure>

The seed holds for the same stock batch, medium, flask and incubation time. Test again when one of them changes.
{% endstep %}
{% endstepper %}

## Related

Working with bacteria that form clumps? [Contact us](https://sbtinstruments.com/contact) to discuss how this works for your organism.


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://help.sbtinstruments.com/inoc/workflows/get-an-active-culture-the-next-day/from-cryo-stock.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.