Replace syringe filter

This page teaches you how to replace the syringe filter on your dilution dispenser.

Replace the syringe filter at the start of each day

Indications of clogged syringe filter

Quick clip

Get things ready

You need these items:

Step by step

1

Uninstall old syringe filter

Unscrew the old syringe filter from the Luer lock adapter on the ejection unit.

2

Install new syringe filter

Screw the new syringe filter onto the Luer lock adapter on the ejection unit.

3

Confirm volume

Dispense once and discard it. The volume is often wrong on the first dispense.

Dispense a second time. Confirm that it fills the vial to the 10 mL mark.

4

That is it!

Remember to replace the syringe filter at the start of each day.

Consult the included Hirschmann manual for comprehensive details on operation, safety, and troubleshooting.