Maintenance schedule
What keeps BactoBox® working well, and how often.
After every measurement, unless the next vial is a less dilute vial of the same sample
Every day bulk dilution liquid is in use
Deep clean (15 minutes), then QC test
Every week BactoBox® is in use, before the first measurement. The QC test page lists the other cases that need one.
Once a year
Every week: deep clean, then QC test
Start each week BactoBox® is in use with a deep clean and a QC test, before the first measurement. Skip both in weeks BactoBox® is not used.
Deep clean for 15 minutes.
QC test. The QC test starts by cleaning with a disinfection vial.
After a period without use, or when BactoBox® is back from service, start with these two steps.
At the end of the day
Run a Clean, power off BactoBox®, and leave the tubing kit in the disinfection vial. Discard that vial the next morning and start the day with a fresh one. For longer periods without use, see Shipping and storage.
Working with a dilution dispenser
With a dilution dispenser, replace the 0.22 µm syringe filter mounted on it at the start of each day. Replace it sooner if it squeaks or needs more force to dispense. Then test the dilution liquid.

See Replace the syringe filter for the full procedure.