Make a QC vial from QC stock
For QC stock, items QC04, QC05 and QC06
You mix QC stock and dilution liquid to make a QC vial. The components are simple but the workflow still requires attention to detail.
Get things ready
You need these items:
1 × SBT ready-made dilution vial
1 × QC stock
1 × vortex mixer (not part of packaging)
1 × P200 pipette (not part of packaging)
Step by step
1
2
Transfer QC stock

Aspirate twice in the QC stock. This provides a more precise volume than a single aspirate.
Use a P200 pipette to transfer 101 µL of QC stock to the dilution vial.
Rinse the pipette tip a few times in the dilution vial, so no QC beads stay on the plastic.
Label the vial “QC”.
3
4
That is it!
You now have a QC vial. Use it to run a QC test.