Run a QC test

Measure a QC vial and a blank to check that BactoBox® measures correctly.

You need these items

Step by step

1

Get a QC vial

Cut out one ready-made QC vial (QC07). Cut out only one at a time.

Cutting one QC vial off the strip

Using QC stock (QC04, QC05 or QC06)? Make a QC vial from QC stock first.

2

Start the QC test program

  1. Press and hold Measure for 2 seconds, until the Programs menu appears.

  2. Press Measure again briefly to select QC test. In this program, Measure works as OK.

  3. Set the target concentration with the buttons. On a ready-made QC vial, it is printed in the center of the sticker. For QC stock, it is on the certificate of analysis.

Setting the target concentration in the QC test program
3

Clean

Move the tubing kit to the disinfection vial and press Measure (OK). If you just ran a Clean, BactoBox® skips this step.

4

Measure the QC vial

Shake the QC vial vigorously for 10 seconds, because the QC beads settle over time. Move the tubing kit to it and press Measure (OK). Press Measure (OK) again to leave the result screen.

Measuring the QC vial
5

Clean again

Move the tubing kit back to the same disinfection vial and press Measure (OK). This rinses out the QC beads.

Cleaning with the disinfection vial
6

Measure the blank

Move the tubing kit to the ready-made dilution vial, with nothing added, and press Measure (OK).

Measuring the dilution vial as the blank
7

Ready for samples

Two check marks on the screen mean the QC test passed. BactoBox® is ready for samples. The QC test also calibrates CIZE for the mounted flow cell.

Discard the QC vial and the dilution vial. Both are for single use. Keep the disinfection vial for the Clean after each sample.

Did the QC test fail?

Run the QC test in Access

If your BactoBox® is connected to Access, run the QC test from your computer. Access guides you through the QC test on screen. Each step unlocks once the previous step completes.

  1. In the bottom bar of Access, click the arrow to the right of Clean and pick QC test.

  2. Follow the four steps on screen. They are the same steps as on your BactoBox®, described above.

  3. At the second step, enter the target concentration in million particles/mL, for example 2.08. Take it from the sticker on a ready-made QC vial, or from the certificate of analysis for QC stock.

The QC test guide in Access.
Access guides you through the QC test, one step at a time.

The pass limits are the same as on your BactoBox®. Access stores each QC test in the QC test measurement group, so you can follow your QC results over time.