Hit the right concentration
Pick the dilution that lands your culture inside the BactoBox® working range.
Your BactoBox® measures accurately when the vial holds 30 000 to 5 000 000 bacteria-sized objects per mL, cells and particles from your medium alike. Most cultures are far denser than that, so you dilute them first.
When you follow a culture through its growth stages, use this page to pick the dilution for each stage.
Pick your dilution
Dilution schemes
Every result inside the working range is accurate. Each dilution scheme on this page aims for about 1 × 10^6 cells/mL in the vial you measure. That gives fast results with little risk of clogging.
1:100 dilutions

1:1 000 dilutions

1:10 000 dilutions

1:100 000 and higher
Extend the 1:10 000 scheme with a 1:10 step. Remove 1 mL of diluent from a third diluent vial, transfer 1 mL from your 1:10 000 vial into it, rinse the pipette tip in the vial, and vortex for 10 seconds. Repeat for each further factor of 10.