Cell size (CIZE)

What CIZE is, and how the QC test keeps it comparable.

CIZE is the average size of the cells your BactoBox® counts. It is given as the diameter of a sphere with the same volume as the cell, in µm. You get it with every measurement, next to cells/mL.

How BactoBox® measures size

Each cell that passes the sensor gives an electrical signal, and the size of that signal scales with the volume of the cell. BactoBox® converts each signal to a diameter. CIZE is the average over all the cells it counts.

Because CIZE is a diameter, it encodes a volume. A CIZE of 2 µm is about eight times the volume of a CIZE of 1 µm.

The QC test calibrates CIZE

The QC vial holds silica beads of a known size. In a QC test, BactoBox® measures the beads and uses their known size to calibrate the size signal of the mounted flow cell. That makes CIZE comparable between devices and between flow cells.

Run a QC test every week you use your BactoBox® to keep CIZE reliable. Your BactoBox® shows a reminder on the home screen when one is due. With ready-made QC vials there is nothing to prepare. When you insert a new flow cell, BactoBox® requires a QC test before it measures, and that test calibrates the new flow cell. See QC test.

CIZE from software older than v2026.09 is hidden

When CIZE reads wrong

Where you see CIZE