Cell size (CIZE)
What CIZE is, and how the QC test keeps it comparable.
CIZE is the average size of the cells your BactoBox® counts. It is given as the diameter of a sphere with the same volume as the cell, in µm. You get it with every measurement, next to cells/mL.
How BactoBox® measures size
Each cell that passes the sensor gives an electrical signal, and the size of that signal scales with the volume of the cell. BactoBox® converts each signal to a diameter. CIZE is the average over all the cells it counts.
Because CIZE is a diameter, it encodes a volume. A CIZE of 2 µm is about eight times the volume of a CIZE of 1 µm.
The QC test calibrates CIZE
The QC vial holds silica beads of a known size. In a QC test, BactoBox® measures the beads and uses their known size to calibrate the size signal of the mounted flow cell. That makes CIZE comparable between devices and between flow cells.
Run a QC test every week you use your BactoBox® to keep CIZE reliable. Your BactoBox® shows a reminder on the home screen when one is due. With ready-made QC vials there is nothing to prepare. When you insert a new flow cell, BactoBox® requires a QC test before it measures, and that test calibrates the new flow cell. See QC test.
When CIZE reads wrong
Clumps. A clump passes the sensor as one object, so it raises CIZE. See Break up cell clumps.
Too few cells. CIZE needs a cells/mL value above the limit of detection.
Where you see CIZE
On the result screen of your BactoBox®.
In the CIZE column in Access. See What the columns mean.
As a dashed line in growth curve plots. See Plot a growth curve.
Was this helpful?