Label your samples
Give each sample its Sampled at time, dilution and label before you measure.
Access stores each result together with the information you give it. With a Sampled at time, a dilution and a label, Access calculates the concentration of your undiluted sample and plots your samples over time. Your BactoBox® cannot record this information on its own.
Fill in the annotation fields
Click the tag icon next to the program buttons in the bottom bar. Four annotation fields appear above the buttons. Fill them in before you press Measure.

Sampled at
The time you took the sample from your culture.
Dilution
The dilution factor of your sample vial. Enter 100 for a 1:100 dilution.
Label
A name for the culture the sample comes from, for example Bioreactor#1.
Measurement group
The group the measurement goes into. See Create and open measurement groups.
Use one label per culture
Access plots one line per label. Give each culture its own label, and use the same label for every sample from that culture. If you run three bioreactors in one measurement group, label them Bioreactor#1, Bioreactor#2 and Bioreactor#3. A label only applies within its measurement group.
Set the time you took the sample
Sampled at decides where the measurement sits on the time axis. Access plots your growth curve from it. Set it to the moment you took the sample from your bioreactor or shake flask, not the moment you measure. Preparing and measuring your sample takes a few minutes, and those minutes are not part of the growth of your culture.
Note the time when you pull the sample. Then set Sampled at to that time in Access, before you press Measure. The clock icon next to the field sets it to now. Use the − and + buttons to adjust the time one minute at a time.
Correct a measurement afterwards
You can change Sampled at, Dilution and Label after the measurement. Open the measurement group, or click Measurements in the side bar. Click the cell, type the new value and press Enter. Then click Save.