Explainer
Linearity across dilutions
Particle type
Tested range (1/mL)
Linearity with dilution factor

Last updated
Was this helpful?
BactoBox® shows great linearity between cells/mL and the dilution factor. This linearity covers the entire range given by the Limit of detection.
The table below shows how BactoBox® performs across various particle types. Note that all R2 values are very close to 1, which indicates strong linearity across the tested range.
Staphyloccus epidermidis
30 000–5 000 000
R2=0.9975
1 µm silica microspheres
1 000 000–5 000 000
R2=0.9995
The below figure illustrates the linearity for the Staphylococcus epidermidis linearity-response test.