For the complete documentation index, see llms.txt. This page is also available as Markdown.
Explainer

Linearity across dilutions

BactoBox® shows great linearity between cells/mL and the dilution factor. This linearity covers the entire range given by the Limit of detection.

The table below shows how BactoBox® performs across various particle types. Note that all R2 values are very close to 1, which indicates strong linearity across the tested range.

Particle type
Tested range (1/mL)
Linearity with dilution factor

Staphyloccus epidermidis

30 000–5 000 000

R2=0.9975

1 µm silica microspheres

1 000 000–5 000 000

R2=0.9995

The below figure illustrates the linearity for the Staphylococcus epidermidis linearity-response test.

S. epidermidis linearity response. Concentration of cells for the same primary sample measured at different dilution factors. Each data point represents arithmetic mean of technical triplicates. Error bars represent standard deviation.

Last updated

Was this helpful?