> For the complete documentation index, see [llms.txt](https://help.sbtinstruments.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://help.sbtinstruments.com/mpd/encyclopedia/cross-check.md).

# Cross-check at end of fermentation

Measure the end of your run with BactoBox® and plate counts. The ratio shows whether to harvest earlier or later.

In this work, we show how BactoBox® can be used with plate counts to tell whether to harvest earlier or later.

For probiotics, vaccines, biostimulants and live biotherapeutics, the product is sold on CFU. Most processes harvest at a fixed time. If that time falls before or after the CFU peak, you lose CFU on every run, and the run itself does not show it.

BactoBox® counts every whole cell. Plate counts count the cells that form a colony. While a culture grows, the two agree. After the peak, cells lose culturability and plate counts fall, while the cell count holds as long as the cells stay whole. One paired measurement at the end of a run therefore shows which side of the peak you are on.

<figure><img src="https://142487182-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FseAjVXi9YrPNSfdp1OXS%2Fuploads%2FEj8F63sBfx2Rs6Cvnni4%2Fcross-check--which-side.png?alt=media" alt="Illustration. Cell count and plate count rise together. After the peak, the plate count falls while the cell count holds. Before the peak, cells and CFU agree, so harvest later. After the peak, cells exceed CFU, so harvest earlier."><figcaption><p>Before the peak, cells and CFU agree. After it, the cell count holds while CFU falls. Illustration.</p></figcaption></figure>

## Take-home messages

* At 13.5 hours, the cell count and the plate count agreed, at a ratio of 1.02. The culture had not yet peaked. Harvesting later would have given about 6× more CFU.
* At 35.8 hours, the cell count was 6.4× the plate count. The culture was past its peak, and CFU was about 7.5× lower than at the peak.
* One paired measurement tells you which way to move your harvest time.

## Study design

* *Pseudomonas fluorescens*, tracked over its full growth curve with BactoBox® and plate counts. Three dilution series per sample for each method.
* We then read two time points as if each were the end of a run, at 13.5 and 35.8 hours.

The [protocol](https://help.sbtinstruments.com/mpd/encyclopedia/cross-check/protocol.md) gives every step.

## Results

<figure><img src="https://142487182-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FseAjVXi9YrPNSfdp1OXS%2Fuploads%2Fr6zs9FS9Jn3HnTHZnsWO%2FPseudomonas_xcheck_introduction.png?alt=media&#x26;token=ecb5a13e-06ac-45d0-91f7-9b62f3e46782" alt=""><figcaption><p><em>P. fluorescens</em> measured with BactoBox® (lavender) and plate counts (yellow). In the cyan window the two agree within 50 %. In the grey window the cell count is well above the plate count. Error bars show the standard deviation of three dilution series.</p></figcaption></figure>

At 13.5 hours the ratio was 1.02 and the error bars overlapped. The plate count peaked later, around 23–29 hours, at about 6× the 13.5-hour value. At 35.8 hours the ratio was 6.40, and the plate count was about 7.5× below its peak.

## Read your own cross-check

We use a 50 % difference as the threshold.

* **Cell count within 50 % of the plate count.** You are before the peak, or right on it. Try a longer incubation.
* **Cell count well above the plate count.** You are past the peak. Try a shorter incubation.
* **Cell count well below the plate count.** Your cells likely clump. Break up the clumps and measure again. See [Break up cell clumps](https://help.sbtinstruments.com/more/how-to-guides/break-up-cell-clumps.md).

Use your cell count to choose the plating dilution. Plate one tenfold dilution on either side of it too, in case the count and the plate count differ.

## What it means for your process

The cross-check costs a few BactoBox® measurements and one set of plate counts. It tells you whether your process has more CFU to give, and in which direction to look. To find the peak itself, [track a growth curve](https://help.sbtinstruments.com/mpd/track-growth-curve.md) with both methods.

## Related

* [Identify harvest time](https://help.sbtinstruments.com/mpd/workflows/mpd1.md), a worked example.
* [Find when growth stops](https://help.sbtinstruments.com/mpd/batch/find-when-growth-stops.md), the workflow.

Questions about your own process? [Contact us](https://sbtinstruments.com/contact).


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://help.sbtinstruments.com/mpd/encyclopedia/cross-check.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.