1:100 dilution after inoculation

Once the shake flask is inoculated, immediately collect a sample to obtain a measurement as close to T0 as possible.

1

Prepare the sample for BactoBox® measurement

  1. Use a serological pipette to transfer a sample from the shake flask to a vial.

  2. Use a permanent marker to label the vial cap with the sampling time.

  3. Return the shake flask to the incubator shaker.

  4. Start a 30-minute countdown to remind you to collect the next sample.

  5. Vortex 30 seconds at max speed to disaggregate bacteria.

  6. Dilute 1:100 in diluent by adding 101 µL of sample to 10 mL of diluent. Vortex 10 sec., max speed.

  7. Transfer the tubing kit to the diluted sample.

Put the flask back in the incubator shaker immediately after collecting the sample.

2

Measure the diluted sample on BactoBox®

  1. Add metadata:

    1. Dilution: 100.

    2. Label: E_coli_growth.

  2. Click the Measure button in Access.

  3. Once measurement is done, click the label E_coli_growth above the chart to visualize the growth curve data.

Do you get the error conductivity too high?

3

Clean the device

  1. Transfer the tubing kit to the disinfection vial.

  2. Click Clean.

Summary

You now know the bacterial concentration at T0. Next we will collect a sample for every 30 min and keep repeating the 1:100 dilution scheme for lag and early exponential growth stage.