# Find when to end the run

See when the culture stops adding cells, and compare it with the titer to decide when to end the run.

At the end of a fed-batch run, the feed continues and the vessel stays busy. That pays only while the culture still adds cells or product.

## Where optical density falls short

Most runs end at a fixed time. Titer, measured afterwards, shows whether the last hours added product. Optical density (OD) often keeps rising to the end, so it cannot show when to stop.

<figure><img src="https://142487182-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FseAjVXi9YrPNSfdp1OXS%2Fuploads%2FavmeEEfA3DwbscWl9U40%2Ff3-end-of-run-v4.svg?alt=media" alt="Illustration in three panels over the feed phase. OD600 keeps rising to the end. Titer peaks, then falls slightly as the feed dilutes the broth. Cells/mL rises in a straight line, keeps rising a little past the titer peak, then falls faster than the feed alone would dilute it. After the titer peak, the run adds feed and vessel time but no product."><figcaption><p>Illustration of real customer data. OD rose to the last sample. The titer peaked, then fell as the feed diluted the broth. The count fell faster than dilution alone, which points to lysis.</p></figcaption></figure>

## Read total cells, not cells/mL

The feed adds volume. Once division stops, cells/mL falls from dilution alone. To see what the cells do, multiply cells/mL by the culture volume. That gives the total cells in the vessel.

* **Total cells flat, cells/mL falling.** Division has stopped, and the feed dilutes the culture.
* **Total cells falling.** Cells are lysing, once clumps are ruled out.

Either way, the culture no longer adds cells.

<figure><img src="https://142487182-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FseAjVXi9YrPNSfdp1OXS%2Fuploads%2FkizpkTWYVnr0esljvzkb%2Ff3-total-cells-v3.svg?alt=media" alt="Illustration in two panels. Left, division has stopped. Total cells stay flat while cells/mL falls as the feed dilutes the culture. Right, cells are lysing. Total cells fall, and cells/mL falls faster."><figcaption><p>Illustration.</p></figcaption></figure>

## Step by step

<figure><img src="https://142487182-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FseAjVXi9YrPNSfdp1OXS%2Fuploads%2FkaOwpM2rDOH4Cpb3XS1b%2Ff3-steps-v4.svg?alt=media" alt="Four steps. 1, record the culture volume at every sample. 2, calculate total cells as cells/mL times volume. 3, find where total cells stop rising, and watch for a fall. 4, compare with the titer. Where titer stops rising, the run adds no new product."><figcaption></figcaption></figure>

{% stepper %}
{% step %}

### Record the volume

Note the culture volume at every sample, and the volume taken out with each sample. Without it, a falling count cannot tell whether the cells lysed or the feed diluted them.
{% endstep %}

{% step %}

### Calculate total cells

Multiply cells/mL by the culture volume at that time, in mL.
{% endstep %}

{% step %}

### Find where total stops rising

Find the point where total cells stop rising. After it, watch for a clear fall.
{% endstep %}

{% step %}

### Compare with the titer

Plot the titer next to total cells. Product can keep forming after the cells stop dividing, so titer decides. Where titer stops rising, the last hours add feed and vessel time, but no product. Decide where to end the run.
{% endstep %}
{% endstepper %}

## The result

A run that ends when the culture stops adding product, read on the cells and the titer, not on OD or the clock.

{% hint style="info" %}
The process owner runs the experiments and makes the decisions. We help read what happens in the culture. [Contact us](https://sbtinstruments.com/contact) to plan a study.
{% endhint %}


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://help.sbtinstruments.com/mpd/fed-batch/find-when-to-end-the-run.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.