Find when to end the run
See when the culture stops adding cells, and compare it with the titer to decide when to end the run.
At the end of a fed-batch run, the feed continues and the vessel stays busy. That pays only while the culture still adds cells or product.
Where optical density falls short
Most runs end at a fixed time. Titer, measured afterwards, shows whether the last hours added product. Optical density (OD) often keeps rising to the end, so it cannot show when to stop.
Read total cells, not cells/mL
The feed adds volume. Once division stops, cells/mL falls from dilution alone. To see what the cells do, multiply cells/mL by the culture volume. That gives the total cells in the vessel.
Total cells flat, cells/mL falling. Division has stopped, and the feed dilutes the culture.
Total cells falling. Cells are lysing, once clumps are ruled out.
Either way, the culture no longer adds cells.
Step by step
The result
A run that ends when the culture stops adding product, read on the cells and the titer, not on OD or the clock.