Protocol
The steps, materials and settings we used to screen the growth media.
This is what we did, step by step, with the materials and settings we used in both studies.
Study 1: four media
What we used
E. coli from one starter culture.
Four media: Terrific broth, LB, TSB and Bacto™ CD Supreme Fermentation Production Medium (CD) with 10 mL/L glycerol. We prepared each medium as its supplier recommends.
One shake flask per medium, for example 50 mL medium in a 250 mL flask.
BactoBox® with a flow cell that had at least 10 measurements left, and a computer with Access.
A photometer for OD600.
Dilution vials, ready to use or filled with the dilution dispenser.
Disinfection vials.
5 mL centrifuge tubes and a rack for 5 mL and 15 mL tubes.
Vortex mixer.
P1000 and P200 pipettes with sterile tips.
Serological pipettes, preferably 2 mL, and a pipette controller.
Marker.
It took 35 hours from inoculation to the last sample. Hands-on time was about 2 hours.
Measurement group
We connected BactoBox® to a computer and opened Access.
We clicked Measurement group in the menu, then Create measurement group. We filled in the details, chose the type Basic and clicked Create. We opened the new group with the open icon. See Create and open measurement groups.
Dilution series
We transferred 101 µL of the sample to 10 mL diluent. We vortexed for 10 s at maximum speed. This gave 1:100.
We transferred 101 µL of the 1:100 dilution to 10 mL diluent. We vortexed for 10 s at maximum speed. This gave 1:10 000.
We expected 1 × 10^10 to 5 × 10^10 cells/mL on the plateau. A 1:10 000 dilution brings that into the working range of BactoBox®.

Measurement
We moved the tubing kit to the 1:10 000 vial. In Access, we filled in these fields:
Dilution: 10 000.
Label: the medium, for example LB or TSB. We used the same label at every time point.
Sampled at: the time we took the sample. The clock icon next to the field sets it to now.
We pressed Measure. Access calculated the concentration in the flask and showed it in Cells/mL × Dil.
We also measured OD600 on each sample with the photometer.
Analysis
The count held on the plateau, so we treated the three time points as replicates for each medium.
In the measurement group, we clicked Export CSV. We opened the file in a spreadsheet and saved it as .xlsx.
For each medium, we calculated the mean and standard deviation of the three time points. In Excel, we used
=AVERAGE()and=STDEV.S().We calculated the coefficient of variation (CV) as the standard deviation divided by the mean.
We ranked the media by their mean cells/mL.
We plotted cells/mL and OD600 at 28 hours side by side in bar charts, to compare the two rankings.

Study 2: glycerol levels in a defined medium
What we used
E. coli ATCC 8739, with the inoculum grown in LB medium.
Bacto™ CD Supreme Fermentation Production Medium, with glycerol added as carbon source.
Three 250 mL baffled Erlenmeyer flasks with vented caps, with 50 mL medium in each, one per glycerol level.
BactoBox® for cells/mL and CIZE.
A photometer for OD600.
Sampling
We sampled each culture through its growth curve. We measured each sample once with BactoBox® and once with OD600. In Access, we set Sampled at to the time the sample left the flask.
For BactoBox®, we dispersed cells (vortex) before the first dilution. Then we diluted the sample into the working range.
Analysis
We plotted cells/mL, OD600 and CIZE against time for each glycerol level.
Questions about your own process? Contact us.
Was this helpful?