For the complete documentation index, see llms.txt. This page is also available as Markdown.

Protocol

The steps, materials and settings we used to screen the growth media.

This is what we did, step by step, with the materials and settings we used in both studies.

Study 1: four media

What we used

Kit for the experiment
The items we used for each round of measurements.
  • E. coli from one starter culture.

  • Four media: Terrific broth, LB, TSB and Bacto™ CD Supreme Fermentation Production Medium (CD) with 10 mL/L glycerol. We prepared each medium as its supplier recommends.

  • One shake flask per medium, for example 50 mL medium in a 250 mL flask.

  • BactoBox® with a flow cell that had at least 10 measurements left, and a computer with Access.

  • A photometer for OD600.

  • Dilution vials, ready to use or filled with the dilution dispenser.

  • Disinfection vials.

  • 5 mL centrifuge tubes and a rack for 5 mL and 15 mL tubes.

  • Vortex mixer.

  • P1000 and P200 pipettes with sterile tips.

  • Serological pipettes, preferably 2 mL, and a pipette controller.

  • Marker.

It took 35 hours from inoculation to the last sample. Hands-on time was about 2 hours.

1

Inoculation

We inoculated one flask of each medium from the same E. coli starter culture. We noted the time. We sampled each flask 22, 28 and 35 hours later, when the cultures were on the plateau.

2

Measurement group

We connected BactoBox® to a computer and opened Access.

We clicked Measurement group in the menu, then Create measurement group. We filled in the details, chose the type Basic and clicked Create. We opened the new group with the open icon. See Create and open measurement groups.

If you are new to Access, see Do measurements in Access.

3

Sampling

At 22 hours, we used a serological pipette to take 1 mL culture from the first flask. We transferred it to a 5 mL centrifuge tube. We noted the time.

4

Cell dispersion

We vortexed the tube for 1 min at maximum speed.

If your strain forms clumps or chains, see Break up cell clumps.

5

Dilution series

  1. We transferred 101 µL of the sample to 10 mL diluent. We vortexed for 10 s at maximum speed. This gave 1:100.

  2. We transferred 101 µL of the 1:100 dilution to 10 mL diluent. We vortexed for 10 s at maximum speed. This gave 1:10 000.

We expected 1 × 10^10 to 5 × 10^10 cells/mL on the plateau. A 1:10 000 dilution brings that into the working range of BactoBox®.

If you work with other organisms, see Hit the right concentration.

Sample workup to a 1:10 000 dilution
The sample workup for a 1:10 000 dilution.
6

Measurement

We moved the tubing kit to the 1:10 000 vial. In Access, we filled in these fields:

  • Dilution: 10 000.

  • Label: the medium, for example LB or TSB. We used the same label at every time point.

  • Sampled at: the time we took the sample. The clock icon next to the field sets it to now.

We pressed Measure. Access calculated the concentration in the flask and showed it in Cells/mL × Dil.

We also measured OD600 on each sample with the photometer.

7

Repeats

We repeated the steps from Sampling to Measurement for the other three media. Then we repeated the whole round at 28 and 35 hours. We ended with one measurement per medium at each time point.

Analysis

The count held on the plateau, so we treated the three time points as replicates for each medium.

  1. In the measurement group, we clicked Export CSV. We opened the file in a spreadsheet and saved it as .xlsx.

  2. For each medium, we calculated the mean and standard deviation of the three time points. In Excel, we used =AVERAGE() and =STDEV.S().

  3. We calculated the coefficient of variation (CV) as the standard deviation divided by the mean.

  4. We ranked the media by their mean cells/mL.

  5. We plotted cells/mL and OD600 at 28 hours side by side in bar charts, to compare the two rankings.

Access data exported to a spreadsheet
Data exported from Access as a CSV file and opened in a spreadsheet.

Check three things before you trust the ranking:

  • A count that rises from the first time point to the later ones means some samples came before the plateau.

  • A CV above 20 % means some samples came before the plateau, or the sample workup varied a lot.

  • Overlapping error bars mean the media reach a similar maximum cell count.

If the result is unclear, repeat the experiment with three measurements per time point. Three flasks per medium is better still.

Study 2: glycerol levels in a defined medium

What we used

  • E. coli ATCC 8739, with the inoculum grown in LB medium.

  • Bacto™ CD Supreme Fermentation Production Medium, with glycerol added as carbon source.

  • Three 250 mL baffled Erlenmeyer flasks with vented caps, with 50 mL medium in each, one per glycerol level.

  • BactoBox® for cells/mL and CIZE.

  • A photometer for OD600.

1

Setup

We added glycerol to the CD medium at 4, 7 and 10 mL/L. The maker of the medium recommends 4–10 mL/L, so 7 mL/L is the center point. We filled 50 mL into each flask.

We inoculated each flask to 1 × 10^7 cells/mL with E. coli ATCC 8739 grown in LB medium.

2

Sampling

We sampled each culture through its growth curve. We measured each sample once with BactoBox® and once with OD600. In Access, we set Sampled at to the time the sample left the flask.

For BactoBox®, we dispersed cells (vortex) before the first dilution. Then we diluted the sample into the working range.

To pick the dilution, see Hit the right concentration.

Analysis

We plotted cells/mL, OD600 and CIZE against time for each glycerol level.

Questions about your own process? Contact us.

Was this helpful?