> For the complete documentation index, see [llms.txt](https://help.sbtinstruments.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://help.sbtinstruments.com/mpd/workflows/check-an-od600-calibration/protocol.md). Agents can also search and read this documentation through its [MCP server](https://help.sbtinstruments.com/~gitbook/mcp).

# Protocol

The steps, materials and settings we used to check an OD600 calibration.

This is what we did, step by step, with the materials and settings we used.

## What we used

* *E. coli* ATCC 8739.
* LB medium.
* One shake flask.
* A shaking incubator at 37 °C and 200 rpm.
* BactoBox® for cells/mL and CIZE.
* An Implen Diluphotometer with standard cuvettes for OD<sub>600</sub>.
* Compact Dry TC plates for plate counts.

Sampling took 11.6 hours.

## What we did

{% stepper %}
{% step %}

### Inoculation

We inoculated LB medium with *E. coli* ATCC 8739 to 1 × 10^6 cells/mL. We incubated the flask at 37 °C and 200 rpm.
{% endstep %}

{% step %}

### Sampling through the growth curve

We took 28 samples over 11.6 hours, starting with the seed. We measured every sample with all three methods, once each. In Access, we set *Sampled at* to the time each sample left the flask.
{% endstep %}

{% step %}

### BactoBox® measurement

We dispersed the cells (vortex) before the first dilution. We diluted the sample into the working range of BactoBox® and measured it once. We noted cells/mL and CIZE.

{% hint style="info" %}
To pick the dilution for your own sample, see [Hit the right concentration](https://help.sbtinstruments.com/more/how-to-guides/hit-the-right-concentration.md).
{% endhint %}
{% endstep %}

{% step %}

### OD<sub>600</sub> measurement

We diluted the sample into the linear range of the photometer. We read OD<sub>600</sub> once in a standard cuvette. We read every sample, also while the culture was too thin to give a signal.

We measured blanks too. We took the detection limit as the blank mean plus three standard deviations.
{% endstep %}

{% step %}

### Plate count

We diluted the sample into the countable range and plated it on a Compact Dry TC plate. We counted the colonies and calculated CFU/mL.
{% endstep %}
{% endstepper %}

## Analysis

For each time point, we calculated three ratios:

* BactoBox® cells/mL per OD unit.
* CFU/mL per OD unit.
* BactoBox® cells/mL per CFU/mL.

We left out the ratios per OD unit where OD<sub>600</sub> was below the detection limit, because dividing by a blank reading divides by noise. In our run, this removed the samples before 2.2 hours.

A ratio is a multiplicative quantity, so we gave each ratio as a geometric mean. We gave its span as the maximum divided by the minimum. For the BactoBox®-to-plate count ratio, we also gave the geometric CV.

We averaged the ratios per OD unit over three growth phases, early exponential (2.2–3.4 h), late exponential (4.4–5.5 h) and stationary (6.1–11.6 h).

To link the drift to cell size, we plotted the ratios per OD unit against CIZE on log–log axes. The slope of a straight-line fit gave the exponent.

Questions about your own process? [Contact us](https://sbtinstruments.com/contact).


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://help.sbtinstruments.com/mpd/workflows/check-an-od600-calibration/protocol.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.