For the complete documentation index, see llms.txt. This page is also available as Markdown.

Protocol

The steps, materials and settings we used to check an OD600 calibration.

This is what we did, step by step, with the materials and settings we used.

What we used

  • E. coli ATCC 8739.

  • LB medium.

  • One shake flask.

  • A shaking incubator at 37 °C and 200 rpm.

  • BactoBox® for cells/mL and CIZE.

  • An Implen Diluphotometer with standard cuvettes for OD600.

  • Compact Dry TC plates for plate counts.

Sampling took 11.6 hours.

What we did

1

Inoculation

We inoculated LB medium with E. coli ATCC 8739 to 1 × 10^6 cells/mL. We incubated the flask at 37 °C and 200 rpm.

2

Sampling through the growth curve

We took 28 samples over 11.6 hours, starting with the seed. We measured every sample with all three methods, once each. In Access, we set Sampled at to the time each sample left the flask.

3

BactoBox® measurement

We dispersed the cells (vortex) before the first dilution. We diluted the sample into the working range of BactoBox® and measured it once. We noted cells/mL and CIZE.

To pick the dilution for your own sample, see Hit the right concentration.

4

OD600 measurement

We diluted the sample into the linear range of the photometer. We read OD600 once in a standard cuvette. We read every sample, also while the culture was too thin to give a signal.

We measured blanks too. We took the detection limit as the blank mean plus three standard deviations.

5

Plate count

We diluted the sample into the countable range and plated it on a Compact Dry TC plate. We counted the colonies and calculated CFU/mL.

Analysis

For each time point, we calculated three ratios:

  • BactoBox® cells/mL per OD unit.

  • CFU/mL per OD unit.

  • BactoBox® cells/mL per CFU/mL.

We left out the ratios per OD unit where OD600 was below the detection limit, because dividing by a blank reading divides by noise. In our run, this removed the samples before 2.2 hours.

A ratio is a multiplicative quantity, so we gave each ratio as a geometric mean. We gave its span as the maximum divided by the minimum. For the BactoBox®-to-plate count ratio, we also gave the geometric CV.

We averaged the ratios per OD unit over three growth phases, early exponential (2.2–3.4 h), late exponential (4.4–5.5 h) and stationary (6.1–11.6 h).

To link the drift to cell size, we plotted the ratios per OD unit against CIZE on log–log axes. The slope of a straight-line fit gave the exponent.

Questions about your own process? Contact us.

Was this helpful?