Protocol
The steps, materials and settings we used to check an OD600 calibration.
This is what we did, step by step, with the materials and settings we used.
What we used
E. coli ATCC 8739.
LB medium.
One shake flask.
A shaking incubator at 37 °C and 200 rpm.
BactoBox® for cells/mL and CIZE.
An Implen Diluphotometer with standard cuvettes for OD600.
Compact Dry TC plates for plate counts.
Sampling took 11.6 hours.
What we did
Analysis
For each time point, we calculated three ratios:
BactoBox® cells/mL per OD unit.
CFU/mL per OD unit.
BactoBox® cells/mL per CFU/mL.
We left out the ratios per OD unit where OD600 was below the detection limit, because dividing by a blank reading divides by noise. In our run, this removed the samples before 2.2 hours.
A ratio is a multiplicative quantity, so we gave each ratio as a geometric mean. We gave its span as the maximum divided by the minimum. For the BactoBox®-to-plate count ratio, we also gave the geometric CV.
We averaged the ratios per OD unit over three growth phases, early exponential (2.2–3.4 h), late exponential (4.4–5.5 h) and stationary (6.1–11.6 h).
To link the drift to cell size, we plotted the ratios per OD unit against CIZE on log–log axes. The slope of a straight-line fit gave the exponent.
Questions about your own process? Contact us.