flask-gearDetermine concentration after incubation

Now, you will determine the concentration after overnight incubation. Cell concentration is typically in a ball-park range of 1 × 1010 cells/mL, and therefore a dilution of 1:10,000 is usually suitable. This dilution is obtained by two sequential 1:100 dilutions. Below we demonstrate this step by step using the Access interface. You can also use BactoBox® as a standalone device and do the calculations manually.

1

Retrieve and disaggregate a sample of the overnight culture

Use a sterile serological pipette to transfer 10 mL of the overnight culture to a sterile vial.

Vortex thoroughly to disperse any cell clumps. For an E. coli culture it is usually sufficient to vortex 30 seconds at max speed.

circle-exclamation

Proper vortexing is important for reliable results!

2

Dilute 1:100

Transfer 101 µL of your sample to 10 mL of diluent. Vortex 10 seconds at maximum speed.

3

Dilute 1:10 000

Transfer 101 µL of the diluted sample to 10 mL of diluent. Vortex 10 seconds at maximum speed.

4

Measure the diluted sample on BactoBox®

Transfer the tubing kit to the 1:10,000 vial.

Press Measure in Access to start a new measurement.

  • Dilution: 10 000

  • Label: Overnight_culture

After completing the measurement, Access will automatically calculate the bacterial concentration in the shake flask. To view the full table, drag the panel sideways.

Summary

Now you know the bacterial concentration of your overnight culture. With cells from an agar plate as input material, you now have a liquid starter culture with a well-defined cell concentration.