1:100 dilutions
The lag phase, acceleration phase, and early exponential growth phase are typically monitored using 1:100 dilutions prepared in 15 mL centrifuge tubes. The procedure for preparing these dilutions is described below.
Special emphasis is placed on proper vortexing of the 15 mL tubes, as inadequate mixing is one of the most common pitfalls in sample preparation for BactoBox® analyses.
How to make 1:100 dilutions
1:100 dilutions are done by transferring 101 µL sample to 10 mL of diluent.

Tutorials 1 explained the importance of using the correct vortexing technique for 15 mL vials. If you need a refresher, expand the section below. Skip ahead if you've already mastered the technique.
Step by step
Repeat 1:100 dilutions for subsequent samples
When the 30-minute timer expires, repeat the same sampling, disaggregation, 1:100 dilution, measurement, and cleaning procedure for the subsequent samples.
Proceed to the 1:1,000 dilutions when the measured cell concentration of the last sample is ≥ 1,500,000 cells/mL.
Summary
You measured the first time point after ~5 hours of incubation. Proceed to 1:1,000 dilutions to determine cell concentration with BactoBox®.
Workarounds to analyze dilute bacterial cultures
To keep this tutorial as simple as possible, only 1:100, 1:1,000, and 1:10,000 dilutions are used. However, custom diluents and dilution factors can also be applied.
Lower dilution factors provide higher sensitivity when measuring dilute bacterial cultures. See Dilute 1:10 for more information.
Last updated
Was this helpful?