BasicTutorial
Deep clean
Get things ready
Step by step
When to deep clean
Want to know more
Last updated
Was this helpful?
You need these items:
1 × BactoBox® setup
1 × deep clean vial
The Deep clean program is done. Discard the deep clean vial. It is for single use only.
Perform a QC test after the Deep clean program is done.
Deep clean your BactoBox® in these cases:
At least once a week. Use the 15-minute option.
To recover a flow cell that failed the QC test. Try with the 15-minute option first. If the issue persists, try the 10-hour option.
See Maintenance for more information.
See Deep clean explained for additional detail.