# Shipping and storage

## Prepare BactoBox® for storage <a href="#ref92370866" id="ref92370866"></a>

The shutdown procedure is simple:

1. Clean the setup (See [Clean](/protocols/best-practices/clean-the-setup.md))
2. Power off the BactoBox® (flip the rocker switch)

Leave the tubing kit in the disinfection vial and keep the flow cell in BactoBox®.

## Prepare BactoBox® for long-term storage <a href="#ref92370866" id="ref92370866"></a>

Step-by-step guide:

1. Clean the setup (See [Clean](/protocols/best-practices/clean-the-setup.md))
2. Power off the BactoBox® (flip the rocker switch)
3. Disconnect [tubing kit](/encyclopedia/item-register/accessories/tubing-kit.md) and place it in an [empty vial](/encyclopedia/item-register/vials-flasks-and-liquids/empty-vial.md) for storage
4. Place the [inlet-outlet cover](/encyclopedia/item-register/accessories/inlet-outlet-cover.md) (the grey protective cap) over BactoBox'® inlet and outlet
5. Preferably store the BactoBox® in the original box together with the [accessories](/encyclopedia/item-register/accessories.md)

{% hint style="success" %}
It is okay (but not mandatory) to leave a flow cell clicked in place during storage.
{% endhint %}

## Prepare BactoBox® for shipping <a href="#toc169268089" id="toc169268089"></a>

Step-by-step guide:

1. Clean the setup (See [Clean](/protocols/best-practices/clean-the-setup.md))
2. Disconnect [tubing kit](/encyclopedia/item-register/accessories/tubing-kit.md) and place it in an [empty vial](/encyclopedia/item-register/vials-flasks-and-liquids/empty-vial.md) for storage
3. Place the [inlet-outlet cover](/encyclopedia/item-register/accessories/inlet-outlet-cover.md) (the grey protective cap) over BactoBox'® inlet and outlet
4. Place the BactoBox® in the original box and then in a padded cardboard box for shipping

{% hint style="success" %}
It is okay (but not mandatory) to leave a flow cell clicked in place during shipping.
{% endhint %}

{% hint style="danger" %}
Do not ship the BactoBox® only in the original box as it does not provide sufficient protection when shipped.
{% endhint %}

## Bring BactoBox® out of storage

Whether you want to bring BactoBox® out of short-term or long-term storage, the procedure is the same: [Qualify your BactoBox® setup](/getting-started/qualification.md).


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://help.sbtinstruments.com/protocols/maintenance/shipping-and-storage.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.