Make sure that you use a clean BactoBox® setup. See Clean.
Step by step
We run the Conductivity program twice (once for each vial). First, a trial run to avoid carry-over from the previous liquid. Then, a final run where we note down the result.
The disinfection vial has zero conductivity. Even a tiny carry-over from the disinfection vial has a large impact on the conductivity result. This is why the trial run is important.
1
Trial run
Transfer the tubing kit to the first vial.
Run the Conductivity program. You find it in the Programs menu.
Press and hold Measure for 1 second from the home screen to access the Programs menu.
Discard the result. This is only a trial run.
2
Final run
Transfer the tubing kit to the second vial.
Run the Conductivity program.
3
That is it!
Use the electrical conductivity from the final run.
Discard both vials.
Your sample vial is for single use only
Once you run the liquid through BactoBox®, the properties change significantly.