BactoBox® software changelog

What we added, changed, or fixed in each software version

We continuously improve the software for your BactoBox®.

Update BactoBox® software

See what changed in each software version in the list below.

Legend

Some changelog entries have icons. The table below describes what each icon means.

Icon
Description

💻

Access, on your computer

📟

The screen and buttons on your BactoBox®

❗

Breaking change May require action or break an existing integration.

Changes

v2026.09

Save your CIZE data before you update

Breaking changes

  • 💻📟❗ Calibrate CIZE with the QC test. CIZE is now comparable between devices and flow cells, and is no longer experimental. Without a calibrating QC test, CIZE stays blank. Run a QC test every week to keep CIZE reliable. See Cell size (CIZE).

  • 💻📟❗ Require a passing QC test on a new flow cell before you can measure. Until then, the device shows "QC test required" and Measure stops with error 2103.

  • 💻❗ Remove the HPI, CFU and OD columns from the basic report and its CSV export.

  • ❗ Drop support for Windows 7.

Added

  • 💻 Connect to Access from a Mac. Access now works on macOS, Windows 10, Windows 11 and Linux.

  • 💻📟 Run the QC test from Access as a guided workflow. The device screen follows the same steps.

  • 💻 Use the new plating, dilution-series and subcultivation calculators. Dilution factors can now be decimals, for example 13.5.

  • 💻 Edit the measurement tables in Access like a spreadsheet, and sort any column.

  • 💻 Download the growth curve, correlation and CIZE plots as SVG files.

  • 💻 Start the 15-minute or the 10-hour deep clean from Access.

Changed

  • 📟 Make the clean program fail when the vial holds a liquid that is not disinfectant, for example dilution liquid.

  • 💻📟 Stop a measurement at start, with a clear message, when a clean is due, the device needs emptying, or the flow cell needs replacing.

  • 📟 Remove the "clean after 20 measurements" reminder. Clean after every measurement instead.

  • 📟 Rework the result screen to show CIZE, and load the History screen faster.

  • 💻 Show numbers the way the device does, whatever your browser language. Decimals use a dot and thousands are separated by spaces.

Fixed

  • 📟 Fix the device getting stuck after a power loss during a program. The device now asks you to empty it.

  • 📟 Fix healthy measurements failing with error 2004 or 2006.

  • 💻📟 Fix measurements losing their result or their error code.

  • 📟 Fix maintenance reminders and device messages that came at the wrong time or gave the wrong advice.

  • Fix device data bundles from v2026.06 that could not be opened. Download a new bundle to replace one.

  • 💻 Fix Access becoming unreachable over USB after a Windows computer wakes from sleep.

  • 💻 Fix glitches in reports, Restore from backup and support cases.

v2026.06a

Fixed

  • 💻 Fix the page buttons in the measurement table. Only the first page of measurements could be reached.

  • 💻 Fix the ID column filter in the measurement table. Filtering by ID made the table fail to load.

v2026.06

Breaking changes

  • ❗ Stop saving a separate .bdr and .csv file for each measurement on the device. The measurements.zip download still holds the .bdr files.

Added

  • 💻 Update your software in one click. Access fetches and installs the latest release.

  • 💻 Start the other programs, for example Conductivity, from a button in Access.

  • 💻 Choose a light, dark or system appearance.

  • 💻 Use keyboard shortcuts, for example F1 to start Measure.

Changed

  • 💻 Rework how Access looks. Everything works as before, but some controls have moved.

  • 💻 Rework the report export.

Fixed

  • Fix a very rare data corruption when you downloaded a device data bundle.

  • 💻 Fix the device data bundle download timing out on the Support page.

  • Show a more specific error message when a measurement fails because objects move too fast.

v2026.02a

Fixed

  • Fix a very rare data corruption.

v2026.02

Added

  • 💻 Add the Growth curve measurement group, with growth plots, cell size and hours post-inoculation.

  • 💻 Filter more columns in the measurement and measurement group tables.

Changed

  • 💻 Rework the Support page. The steps to create a support case are the same.

Fixed

  • Fix the "Database is locked" error when you connect to Access while a program runs.

v2026.01

Added

  • 💻 View and label your measurements on the new Measurements page.

  • 💻 Collect measurements in measurement groups, with growth curve and CFU correlation plots.

  • 💻 Start and stop programs from Access with the Measure, Stop, Empty and Clean buttons.

  • 💻 Follow the program that runs in a progress bar.

  • 💻 Set the clock of your BactoBox® automatically when you connect to Access.

  • 💻 See your system info at the top right of Access.

  • 💻 Open the help center from an error code in Access.

Changed

  • Write "mL" instead of "ml" everywhere.

  • 💻 Rename the Summary page to Download summary.

Fixed

  • 📟 Fix the screen getting stuck when a program fails to start.

  • Fix measurements where fields such as Software version were empty.

  • Fix device data bundles that could be damaged.

  • 💻 Fix the layout of Access on small screens.

v2025.10b

Changed

  • 📟 Rename "blank diluent" to "Dilution vial" in the QC test.

Fixed

  • Fix an error when you pressed Measure after a gating file was damaged.

  • Save gating changes at once, so they are kept if you switch off right after you save.

v2025.10

Added

  • Choose a 15-minute or a 10-hour Deep clean.

  • 📟 Add the steps Clean and Measure on blank diluent to the QC test on the screen.

  • 📟 Add a Set target concentration screen before the QC test.

  • Add the 50XX error codes for the blank liquid check in the QC test.

Changed

  • Point error messages to help.sbtinstruments.com.

  • 📟 Enable Empty only when there is liquid in the device.

  • 📟 Show only Ready on the home screen.

  • Replace the Flush program with the 15-minute Deep clean.

  • Remind you to deep clean when a week has passed since the last deep clean.

  • Use "Clean the setup" in device messages, to match the help center.

  • Stop counting the QC test as a measurement on the flow cell.