For the complete documentation index, see llms.txt. This page is also available as Markdown.
Reference

Known software issues

Some software issues deserve special mention. This is what this page is for.

Unfixed issues

baxter-oom

Trigger

Triggers during a measurement.

Symptoms

All subsequent measurements fail with:

[Errno 111] Connection refused

Recovery

Restart your BactoBox® to recover.

Metadata

🪲 Introduced in initial version

🔧 Fixed in N/A (not fixed yet)

🎲 Likelihood is Rare. Approximately 1 in 1000 measurements.

🌡️ Severity is Low. No data loss.

Fixed issues

wal-db-corruption

Trigger

Triggers after first connection to Access.

Symptoms

Symptoms include:

  • Data loss (rare but this is the worst case scenario)

  • Unable to start any program (even after reboot)

Recovery

First, try to restart your BactoBox®.

If you are unable to start any program, delete all user data and then restart your BactoBox®.

Metadata

🪲 Introduced in v2026.02-alpha.0

🔧 Fixed in v2026.02a . We accidentally released a POSIX advisory lock on the database.

🎲 Likelihood is Rare

🌡️ Severity is High. Worst case is complete data loss.

sudden-system-failure

Trigger

Triggers at any point in time.

Symptoms

One or more of the following happens:

  • Screen hangs (freezes)

  • Screen glitches

  • Software becomes completely unresponsive to further user input (button presses)

  • Device powers itself off

  • Various error messages (e.g., "no response")

Recovery

Restart your BactoBox® to recover.

Metadata

🪲 Introduced in initial version

🔧 Fixed in v2024.03 . Was a bug in the spi-xilinx linux kernel driver. We sent a patch to Xilinx (now AMD).

🎲 Likelihood is Rare. Once every 100 hours if left powered on at idle. Slightly more frequent if device is under constant use.

🌡️ Severity is Low. No data loss.

Last updated

Was this helpful?