For the complete documentation index, see llms.txt. This page is also available as Markdown.

BactoBox® configuration

BactoBox® comes with a factory-defined configuration. A configuration is a set of default settings. It covers About gating and result presentation.

Configuration is independent of the hardware version and software version.

Current configurations

Nowadays, there is a single BactoBox® configuration: BB14 (aka "Customizable"). We recommend BB14 for all new users.

The configuration name matches the item number used to purchase BactoBox®.

Legacy configurations

Configuration
Also known as
Notes

BB12

EasyCount

No longer available.

BB05

Original

No longer available. For adopters of an early version of BactoBox®.

Configuration-defined defaults

These are defaults. You can sometimes override them.

Example: you can select a different CRS file.

Configuration
Gating (CRS file)
Limit of detection
Presentation

BB14 Customizable

BacTotal_v2024-10 You can change this.

core2b-latest 30 000–5 000 000/mL

sbt2_v2025-01 Example: "total/mL"

BB12 EasyCount No longer available

BacTotal_v2024-10 You cannot change this.

core2b-latest 30 000–5 000 000/mL

sbt2_v2025-01 Example: "total/mL"

BB05 Original No longer available

BacTotal_v2024-02 You can change this.

sbt1-latest 5 000–5 000 000/mL

sbt1_v2022-03 Example: "total particles/mL"

Was this helpful?