For the complete documentation index, see llms.txt. This page is also available as Markdown.

Technical specification

Measurement

Parameter
Value

Measurement principle

Electrical impedance flow cytometry

Target organisms

Bacterial cells in suspension

Output parameters

Cell concentration (cells/mL), cell size

Linear quantification range

3 × 10⁴ – 5 × 10⁶ cells/mL

Measurement time

~2 minutes per sample

Sample preparation

Dilution required. Homogenization needed for aggregating samples.

Precision and accuracy

Instrument

Parameter
Value

Form factor

Benchtop instrument

Dimensions (W × D × H)

240 × 140 × 60 mm (9.4 × 5.5 × 2.4 in)

Net weight

1 200 g

Input voltage

12 V⎓ / 2 500 mA / 30 W

Software

Included. Visualization, growth curves, and data export.

Operating conditions

Parameter
Value

Operating temperature

20–25 °C

Extended operating temperature

10–40 °C. Contact SBT first.

Power usage

BactoBox® power usage in various situations.

Situation
Average
Peak

Idle

3.60 W

3.84 W

Measure

6.00 W

13.9 W

The power usage peaks whenever the flow changes direction or quickly accelerates. These peaks usually last for a very short duration.

Was this helpful?