> For the complete documentation index, see [llms.txt](https://help.sbtinstruments.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://help.sbtinstruments.com/troubleshooting/error-code-on-screen/qc-test-fails/blank-fails.md).

# Blank fails

Codes 5010 and 5011. Clean with a fresh disinfection vial, then measure a fresh dilution vial.

<figure><img src="https://1589625940-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FQorx08wID6YNpw8pEV0j%2Fuploads%2F3Lav7oKI3LW8FpR8FrVb%2Fblank-fails.svg?alt=media" alt="The blank fails with code 5010 or 5011. Replace the used disinfection vial and clean with a fresh one. Then measure a fresh dilution vial as the blank. It passes."><figcaption></figcaption></figure>

A blank most often fails right after a deep clean. The deep clean loosens buildup in the flow path, and the Clean that follows flushes it into the disinfection vial. When the QC test cleans again before the blank with that same vial, the particles are drawn back in and counted in the blank.

A Clean with a fresh disinfection vial removes them. A fresh dilution vial rules out the vial itself.

## Step by step

{% stepper %}
{% step %}

#### Get fresh vials

* 1 × fresh SBT [disinfection vial](https://help.sbtinstruments.com/item-register/consumables/disinfection-vial.md)
* 1 × fresh SBT ready-made [dilution vial](https://help.sbtinstruments.com/item-register/individual-items/dilution-vial.md)

Discard the used disinfection vial.
{% endstep %}

{% step %}

#### Clean with the fresh disinfection vial

The QC test program goes back to the *Clean* step. Move the tubing kit to the fresh disinfection vial and press <kbd>Measure</kbd> (<kbd>OK</kbd>).
{% endstep %}

{% step %}

#### Measure the fresh dilution vial

Follow the screen. Use the fresh dilution vial for the blank. If the program asks for the QC vial again, use a fresh QC vial too.
{% endstep %}

{% step %}

#### Still failing?

Go to [Check your consumables](https://help.sbtinstruments.com/troubleshooting/error-code-on-screen/qc-test-fails/check-your-consumables.md).
{% endstep %}
{% endstepper %}


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://help.sbtinstruments.com/troubleshooting/error-code-on-screen/qc-test-fails/blank-fails.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.