# Error codes

Look up the code on your BactoBox® screen and follow the steps.

Find your code in the table. Use the column for where the error appeared, during a sample or during a QC test. Follow the steps in order, and stop as soon as the error is gone. If the last step does not fix it, [contact support](/troubleshooting/contact-support.md).

Two steps come up often.

* **Clean and try again.** Run [Clean](/more/how-to-guides/clean-your-setup.md) with a fresh SBT [disinfection vial](/item-register/consumables/disinfection-vial.md), even if your current one has uses left. Then measure a new sample vial, or run the QC test again with a fresh SBT QC vial and a fresh SBT ready-made dilution vial.
* **Restart.** Switch BactoBox® off and on with the rocker switch. Wait for the *Home* screen, then try again with fresh vials.

{% hint style="info" %}

### When a step says *Run a QC test*

Run it with SBT consumables only, so the result tells you about your BactoBox®. The result decides what comes next. If the QC test passes, your BactoBox® works and the cause is in the sample. See [Sample gives an error](/troubleshooting/sample-gives-an-error.md). If it fails, follow the QC test column for the code it shows.
{% endhint %}

{% hint style="success" %}

### Use fresh vials at room temperature

Alcohol left inside BactoBox® after Clean can lower the conductivity, so a vial you have already measured may read too low. Warm or cold liquid also changes the conductivity. Try again with a fresh vial, and make sure your dilution liquid, QC vial and dilution vial are at room temperature.
{% endhint %}

## Codes

| Code | Message                                              | During a sample                                                                                                                                                                                                 | During a QC test                                                                                                                                                                                 |
| ---- | ---------------------------------------------------- | --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| 1000 | Electrical signal is too low                         | Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                                                                      | [Remove and reinsert the flow cell](/getting-started/installation/install-flow-cell.md), then clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md) |
| 1001 | Electrical signal is too high                        | Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                                                                      | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 1101 | Electrical noise is too high                         | Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                                                                      | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 1111 | Electrical noise is too high                         | Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                                                                      | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 1121 | Electrical noise is too high                         | Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                                                                      | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 1131 | Electrical noise is too high                         | Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                                                                      | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 1200 | Conductivity is too low…                             | Clean and try again → Check that you use SBT dilution liquid at room temperature → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                         | Check that the QC vial and dilution vial are at room temperature → Clean and try again → [Check liquid uptake](/troubleshooting/qc-test-fails/check-liquid-uptake.md)                            |
| 1201 | Conductivity is too high…                            | Check that your dilution liquid is at room temperature, not warm → Dilute more, for example 1:1 000, to see if the growth medium raises the conductivity → [Run a QC test](/maintenance/qc-test/run-qc-test.md) | Check that the QC vial and dilution vial are at room temperature → Restart                                                                                                                       |
| 1202 | Conductivity was not stable…                         | Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                                                                      | Check that the QC vial and dilution vial are at room temperature → Clean and try again → Restart                                                                                                 |
| 1203 | Conductivity sensor is malfunctioning                | Clean and try again → Restart                                                                                                                                                                                   | Clean and try again → Restart                                                                                                                                                                    |
| 1204 | Could not read from conductivity sensor              | Clean and try again → Restart                                                                                                                                                                                   | Clean and try again → Restart                                                                                                                                                                    |
| 1205 | Conductivity is way too high…                        | Check that your dilution liquid is at room temperature, not warm → Dilute more, for example 1:1 000, to see if the growth medium raises the conductivity → [Run a QC test](/maintenance/qc-test/run-qc-test.md) | Check that the QC vial and dilution vial are at room temperature → Restart                                                                                                                       |
| 1206 | Conductivity is way too low…                         | Clean and try again → Check that you use SBT dilution liquid at room temperature → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                         | Check that the QC vial and dilution vial are at room temperature → Clean and try again → [Check liquid uptake](/troubleshooting/qc-test-fails/check-liquid-uptake.md)                            |
| 1207 | Cleaning liquid conductivity is too high…            | Appears during *Clean*. Clean again with a fresh SBT [disinfection vial](/item-register/consumables/disinfection-vial.md) at room temperature.                                                                  | Appears during *Clean*. Clean again with a fresh SBT [disinfection vial](/item-register/consumables/disinfection-vial.md) at room temperature.                                                   |
| 2000 | Detected events are too fast                         | If your sample holds many particles, dilute it more. Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                 | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 2001 | Detected events are too slow                         | If your sample holds many particles, dilute it more. Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                 | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 2004 | Liquid moves slower than expected                    | If your sample holds many particles, dilute it more. Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                 | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 2006 | Flow did not stabilize in time                       | If your sample holds many particles, dilute it more. Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                 | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 2010 | Velocity differs between detection sites             | If your sample holds many particles, dilute it more. Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                 | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 2020 | Concentration differs between detection sites        | If your sample holds many particles, dilute it more. Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                 | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 2030 | Flow cell may be clogged                             | If your sample holds many particles, dilute it more. Clean and try again → [Run a QC test](/maintenance/qc-test/run-qc-test.md)                                                                                 | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 2103 | QC test required                                     | Appears when you press *Measure* on a new flow cell. [Run a QC test](/maintenance/qc-test/run-qc-test.md).                                                                                                      | Does not occur                                                                                                                                                                                   |
| 3000 | Particle concentration is too high…                  | Dilute more. Clean and try again. See [Hit the right concentration](/more/how-to-guides/hit-the-right-concentration.md).                                                                                        | Does not occur                                                                                                                                                                                   |
| 4001 | No objects detected                                  | Does not occur                                                                                                                                                                                                  | Restart → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                                           |
| 4010 | …total cells/mL is outside the allowed range…        | Does not occur                                                                                                                                                                                                  | Clean and try again → [Check your consumables](/troubleshooting/qc-test-fails/check-your-consumables.md)                                                                                         |
| 4011 | …particles/mL is outside the allowed range…          | Does not occur                                                                                                                                                                                                  | Clean and try again → [Check your consumables](/troubleshooting/qc-test-fails/check-your-consumables.md)                                                                                         |
| 4012 | Detected object velocities are outside normal limits | Does not occur                                                                                                                                                                                                  | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 4013 | HF phase is not within…                              | Does not occur                                                                                                                                                                                                  | [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                                                     |
| 4014 | HF amplitude is not within…                          | Does not occur                                                                                                                                                                                                  | [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                                                     |
| 4020 | QC test failed…                                      | Does not occur                                                                                                                                                                                                  | Clean and try again → [Check the flow cell](/troubleshooting/qc-test-fails/check-the-flow-cell.md)                                                                                               |
| 5010 | …total cells/mL is outside the allowed range…        | Does not occur                                                                                                                                                                                                  | Clean and try again → [Check your consumables](/troubleshooting/qc-test-fails/check-your-consumables.md)                                                                                         |
| 5011 | …particles/mL is outside the allowed range…          | Does not occur                                                                                                                                                                                                  | Clean and try again → [Check your consumables](/troubleshooting/qc-test-fails/check-your-consumables.md)                                                                                         |
| 8100 | Measurement did not compute final data in time       | Restart → [Update BactoBox® software](/more/software/update-bactobox-r-software.md)                                                                                                                             | Restart → [Update BactoBox® software](/more/software/update-bactobox-r-software.md)                                                                                                              |
| 9969 | The electronics may be damaged                       | Restart → [Update BactoBox® software](/more/software/update-bactobox-r-software.md)                                                                                                                             | Restart → [Update BactoBox® software](/more/software/update-bactobox-r-software.md)                                                                                                              |
| 9970 | Could not read reference file                        | Restart → [Update BactoBox® software](/more/software/update-bactobox-r-software.md)                                                                                                                             | Restart → [Update BactoBox® software](/more/software/update-bactobox-r-software.md)                                                                                                              |

In a QC test, codes 4001 to 4020 come from the QC vial. Codes 5010 and 5011 come from the blank, the plain dilution vial measured at the end.

The wording of a message can differ between software versions. The code stays the same. Code not listed? [Contact support](/troubleshooting/contact-support.md).

## Messages without a code

| Message                                                                      | Go to                                                                                                         |
| ---------------------------------------------------------------------------- | ------------------------------------------------------------------------------------------------------------- |
| Program could not run to completion                                          | [Program could not run to completion](/troubleshooting/other-problems/program-could-not-run-to-completion.md) |
| Flow cell invalid, flow cell obsolete, flow cell depleted, flow cell removed | [Flow cell does not work](/troubleshooting/other-problems/flow-cell-does-not-work.md)                         |
| Update failed                                                                | [Software update failed](/troubleshooting/other-problems/software-update-failed.md)                           |
| This site can't be reached                                                   | [Cannot connect to Access](/troubleshooting/other-problems/cannot-connect-to-access.md)                       |
| Initializing, Stopping                                                       | [Stuck at spinner](/troubleshooting/other-problems/stuck-at-spinner.md)                                       |

For other problems without a message, see [Other problems](/troubleshooting/other-problems.md).


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://help.sbtinstruments.com/troubleshooting/error-codes.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.