Error codes
Look up the code on your BactoBox® screen and follow the steps.
Find your code in the table. Use the column for where the error appeared, during a sample or during a QC test. Follow the steps in order, and stop as soon as the error is gone. If the last step does not fix it, contact support.
Two steps come up often.
Clean and try again. Run Clean with a fresh SBT disinfection vial, even if your current one has uses left. Then measure a new sample vial, or run the QC test again with a fresh SBT QC vial and a fresh SBT ready-made dilution vial.
Restart. Switch BactoBox® off and on with the rocker switch. Wait for the Home screen, then try again.
When a step says Run a QC test
Run it with SBT consumables only, so the result tells you about your BactoBox®. The result decides what comes next. If the QC test passes, your BactoBox® works and the cause is in the sample. See Sample gives an error. If it fails, follow the QC test column for the code it shows.
Codes
1000
Electrical signal is too low
Clean and try again → Run a QC test
Remove and reinsert the flow cell, then clean and try again → Check the flow cell
1001
Electrical signal is too high
Clean and try again → Run a QC test
Clean and try again → Check the flow cell
1101
Electrical noise is too high
Clean and try again → Run a QC test
Clean and try again → Check the flow cell
1111
Electrical noise is too high
Clean and try again → Run a QC test
Clean and try again → Check the flow cell
1121
Electrical noise is too high
Clean and try again → Run a QC test
Clean and try again → Check the flow cell
1131
Electrical noise is too high
Clean and try again → Run a QC test
Clean and try again → Check the flow cell
1200
Conductivity is too low…
Clean and try again → Check that you use SBT dilution liquid at room temperature → Run a QC test
Check that the QC vial and dilution vial are at room temperature → Clean and try again → Check liquid uptake
1201
Conductivity is too high…
Check that your dilution liquid is at room temperature, not warm → Dilute more, for example 1:1 000, to see if the growth medium raises the conductivity → Run a QC test
Check that the QC vial and dilution vial are at room temperature → Restart
1202
Conductivity was not stable…
Clean and try again → Run a QC test
Check that the QC vial and dilution vial are at room temperature → Clean and try again → Restart
1203
Conductivity sensor is malfunctioning
Clean and try again → Restart
Clean and try again → Restart
1204
Could not read from conductivity sensor
Clean and try again → Restart
Clean and try again → Restart
1205
Conductivity is way too high…
Check that your dilution liquid is at room temperature, not warm → Dilute more, for example 1:1 000, to see if the growth medium raises the conductivity → Run a QC test
Check that the QC vial and dilution vial are at room temperature → Restart
1206
Conductivity is way too low…
Clean and try again → Check that you use SBT dilution liquid at room temperature → Run a QC test
Check that the QC vial and dilution vial are at room temperature → Clean and try again → Check liquid uptake
1207
Cleaning liquid conductivity is too high…
Appears during Clean. Clean again with a fresh SBT disinfection vial at room temperature.
Appears during Clean. Clean again with a fresh SBT disinfection vial at room temperature.
2000
Detected events are too fast
If your sample holds many particles, dilute it more. Clean and try again → Run a QC test
Clean and try again → Check the flow cell
2001
Detected events are too slow
If your sample holds many particles, dilute it more. Clean and try again → Run a QC test
Clean and try again → Check the flow cell
2004
Liquid moves slower than expected
If your sample holds many particles, dilute it more. Clean and try again → Run a QC test
Clean and try again → Check the flow cell
2006
Flow did not stabilize in time
If your sample holds many particles, dilute it more. Clean and try again → Run a QC test
Clean and try again → Check the flow cell
2010
Velocity differs between detection sites
If your sample holds many particles, dilute it more. Clean and try again → Run a QC test
Clean and try again → Check the flow cell
2020
Concentration differs between detection sites
If your sample holds many particles, dilute it more. Clean and try again → Run a QC test
Clean and try again → Check the flow cell
2030
Flow cell may be clogged
If your sample holds many particles, dilute it more. Clean and try again → Run a QC test
Clean and try again → Check the flow cell
3000
Particle concentration is too high…
Dilute more. Clean and try again. See Hit the right concentration.
Does not occur
4010
…total cells/mL is outside the allowed range…
Does not occur
Clean and try again → Check your consumables
4011
…particles/mL is outside the allowed range…
Does not occur
Clean and try again → Check your consumables
4012
Detected object velocities are outside normal limits
Does not occur
Clean and try again → Check the flow cell
5010
…total cells/mL is outside the allowed range…
Does not occur
Clean and try again → Check your consumables
5011
…particles/mL is outside the allowed range…
Does not occur
Clean and try again → Check your consumables
8100
Measurement did not compute final data in time
Restart → Update BactoBox® software
Restart → Update BactoBox® software
9969
The electronics may be damaged
Restart → Update BactoBox® software
Restart → Update BactoBox® software
9970
Could not read reference file
Restart → Update BactoBox® software
Restart → Update BactoBox® software
In a QC test, codes 4001 to 4020 come from the QC vial. Codes 5010 and 5011 come from the blank, the plain dilution vial measured at the end.
The wording of a message can differ between software versions. The code stays the same. Code not listed? Contact support.
Messages without a code
Program could not run to completion
Flow cell invalid, flow cell obsolete, flow cell depleted, flow cell removed
Update failed
This site can't be reached
Initializing, Stopping
For other problems without a message, see Other problems.