Dilution liquid test fails
Your blank dilution liquid reads above the limits. Find the cause before you test again.
Your dilution liquid without any added medium or cells read above 30 000 cells/mL or above 100 000 total/mL. Do not just test again. Work through these steps in order, and stop when the test passes.
Use only SBT items
When the filter is not the cause, an item from another supplier usually is.
Vials. Use SBT empty vials, not your own 15 mL centrifuge tubes. Tubes from some manufacturers release particles into the liquid.
Disinfection liquid. Use SBT disinfection vials, not your own 70 % isopropanol or ethanol. We have seen such liquids add particles that carry over into the next measurement.
Replace any item that is not from SBT.
Test again
Have an SBT ready-made dilution vial? Measure it first, then run Clean.
Fill a new SBT empty vial with your dilution liquid and measure it. Do not measure the vial that failed again.
If the ready-made vial passes and your filled vial fails, the cause is in your liquid, dispenser, filter or vial.
Still failing? Contact us.