Dilution liquid test fails

Your blank dilution liquid reads above the limits. Find the cause before you test again.

Your dilution liquid without any added medium or cells read above 30 000 cells/mL or above 100 000 total/mL. Do not just test again. Work through these steps in order, and stop when the test passes.

1

Replace the filter on the dilution dispenser

A used or faulty filter is the most common cause. Replace the 0.22 µm syringe filter mounted on the dilution dispenser, even if you replaced it this morning.

2

Use only SBT items

When the filter is not the cause, an item from another supplier usually is.

  • Vials. Use SBT empty vials, not your own 15 mL centrifuge tubes. Tubes from some manufacturers release particles into the liquid.

  • Disinfection liquid. Use SBT disinfection vials, not your own 70 % isopropanol or ethanol. We have seen such liquids add particles that carry over into the next measurement.

Replace any item that is not from SBT.

3

Fill vials on the day you use them

Do not fill vials the day before. Bacteria grow in a filled vial overnight.

Do not store filled vials in a fridge either. Cold storage changes the conductivity, and your BactoBox® will likely show errors.

4

Clean with a fresh disinfection vial

Run Clean with a new SBT disinfection vial. A disinfection vial you have used for other things can carry particles over into the test.

5

Test again

Have an SBT ready-made dilution vial? Measure it first, then run Clean.

Fill a new SBT empty vial with your dilution liquid and measure it. Do not measure the vial that failed again.

If the ready-made vial passes and your filled vial fails, the cause is in your liquid, dispenser, filter or vial.

Still failing? Contact us.