Flow cell does not work

The screen says the flow cell is invalid, obsolete, depleted or removed.

Find the message your BactoBox® shows.

Got an error code during a sample or a QC test instead? See Error codes.

Flow cell removed

Your BactoBox® does not detect the flow cell.

Press firmly on the top section of the flow cell until it clicks. Then press firmly on the lower section until it clicks again. Once in place, it feels slightly springy when you press the center. See Install flow cell.

Press the flow cell into place, top section first, then the lower section.

If the message stays, contact support.

Flow cell invalid

Remove the flow cell and install it again. If the message stays, contact support.

Flow cell depleted

The flow cell has reached 250 measurements and is locked. The limit protects its factory calibration. Install a new flow cell, and replace the external filter at the same time.

Flow cell obsolete

This flow cell is an old type that your BactoBox® no longer supports. Install a current flow cell.