Blank fails
Codes 5010 and 5011. Clean with a fresh disinfection vial, then measure a fresh dilution vial.
A blank most often fails right after a deep clean. The deep clean loosens buildup in the flow path, and the Clean that follows flushes it into the disinfection vial. When the QC test cleans again before the blank with that same vial, the particles are drawn back in and counted in the blank.
A Clean with a fresh disinfection vial removes them. A fresh dilution vial rules out the vial itself.
Step by step
1
Get fresh vials
1 × fresh SBT disinfection vial
1 × fresh SBT ready-made dilution vial
Discard the used disinfection vial.
2
3
4
Still failing?
Go to Check your consumables.