Blank fails

Codes 5010 and 5011. Clean with a fresh disinfection vial, then measure a fresh dilution vial.

The blank fails with code 5010 or 5011. Replace the used disinfection vial and clean with a fresh one. Then measure a fresh dilution vial as the blank. It passes.

A blank most often fails right after a deep clean. The deep clean loosens buildup in the flow path, and the Clean that follows flushes it into the disinfection vial. When the QC test cleans again before the blank with that same vial, the particles are drawn back in and counted in the blank.

A Clean with a fresh disinfection vial removes them. A fresh dilution vial rules out the vial itself.

Step by step

1

Get fresh vials

Discard the used disinfection vial.

2

Clean with the fresh disinfection vial

The QC test program goes back to the Clean step. Move the tubing kit to the fresh disinfection vial and press Measure (OK).

3

Measure the fresh dilution vial

Follow the screen. Use the fresh dilution vial for the blank. If the program asks for the QC vial again, use a fresh QC vial too.

4

Still failing?