Check liquid uptake

Make sure your BactoBox® draws liquid from the vial. If not, replace the parts in the flow path one at a time.

Your BactoBox® measures the conductivity of the liquid it draws in. If no liquid comes in, the conductivity reads too low. Then a part of the flow path is most likely clogged. Replace the parts one at a time, so you know which one was the cause. If the QC test fails with another code along the way, follow that code in Error codes.

Step by step

1

Watch the vial

Run a QC test again with fresh vials at room temperature. When the measurement starts, watch the liquid level in the vial. It should drop.

  • The level drops. Liquid reaches your BactoBox®. Contact support.

  • The level does not drop. Go on to the next step.

2

Replace the external filter

Replace the external filter, as shown in Install tubing kit. Run a QC test. If it passes, you are done.

3

Replace the flow cell

Install a new flow cell. Keep the old one, so you can try to recover it later. Run a QC test. If it passes, you are done.

4

Replace the tubing kit

Install a new tubing kit. You can order a spare tubing kit from SBT. Run a QC test. If it passes, you are done.

5

Contact support

QC test still fails? Contact support. Tell us which parts you replaced.