Check liquid uptake
Make sure your BactoBox® draws liquid from the vial. If not, replace the parts in the flow path one at a time.
Your BactoBox® measures the conductivity of the liquid it draws in. If no liquid comes in, the conductivity reads too low. Then a part of the flow path is most likely clogged. Replace the parts one at a time, so you know which one was the cause. If the QC test fails with another code along the way, follow that code in Error codes.
Step by step
Watch the vial
Run a QC test again with fresh vials at room temperature. When the measurement starts, watch the liquid level in the vial. It should drop.
The level drops. Liquid reaches your BactoBox®. Contact support.
The level does not drop. Go on to the next step.
Replace the external filter
Replace the external filter, as shown in Install tubing kit. Run a QC test. If it passes, you are done.
Replace the flow cell
Install a new flow cell. Keep the old one, so you can try to recover it later. Run a QC test. If it passes, you are done.
Replace the tubing kit
Install a new tubing kit. You can order a spare tubing kit from SBT. Run a QC test. If it passes, you are done.
Contact support
QC test still fails? Contact support. Tell us which parts you replaced.