Check the flow cell

Look for liquid under the flow cell. Then try a new flow cell and external filter.

A QC test error can come from the flow cell, or from how your samples have affected the flow cell. A new flow cell tells you if the old one is the cause.

Get things ready

You need these items:

A QC test does not use any of the 250 measurements on a flow cell.

Step by step

1

Clean

Run Clean before you remove the flow cell.

2

Remove the flow cell

Remove the flow cell cover. Press the clamps together and lift the flow cell out. Keep it, so you can try to recover it later.

3

Look for liquid

Look into the flow cell compartment for liquid or drops.

Found liquid? Stop here

4

Install a new flow cell and external filter

Install the new flow cell. Then replace the external filter, as shown in Install tubing kit.

5

Run a QC test

Run a QC test with fresh vials.

After error 1000, do not try to recover the old flow cell

A deep clean does not fix error 1000. If the old flow cell is still under warranty, contact support. See the warranty on the flow cell page.