Check the flow cell
Look for liquid under the flow cell. Then try a new flow cell and external filter.
A QC test error can come from the flow cell, or from how your samples have affected the flow cell. A new flow cell tells you if the old one is the cause.
Get things ready
You need these items:
1 × new flow cell
1 × new external filter. Every flow cell ships with two.
Everything you need to run a QC test
Step by step
Clean
Run Clean before you remove the flow cell.
Remove the flow cell
Remove the flow cell cover. Press the clamps together and lift the flow cell out. Keep it, so you can try to recover it later.
Install a new flow cell and external filter
Install the new flow cell. Then replace the external filter, as shown in Install tubing kit.
Run a QC test
Run a QC test with fresh vials.
QC test passes. The old flow cell was the cause. Keep using the new one. When you have time, try to recover the old flow cell.
QC test fails. Contact support.
After error 1000, do not try to recover the old flow cell
A deep clean does not fix error 1000. If the old flow cell is still under warranty, contact support. See the warranty on the flow cell page.