For the complete documentation index, see llms.txt. This page is also available as Markdown.

Check your consumables

Use SBT consumables for all your measurements. When you check if your BactoBox® works, use SBT consumables only.

We recommend SBT consumables for all your measurements, samples included. They rule out many errors that are very hard for us to trace in a sample.

When you check if your BactoBox® works, run every QC test with SBT consumables only. Then the result tells you about your BactoBox®, not about a vial or a liquid. If the QC test passes, a vial or liquid you used before was the cause. If it fails, the cause is in your BactoBox®.

Step by step

1

Check the target concentration

Did the QC vial fail, with code 4010 or 4011? Compare the target concentration you entered with the certificate of analysis for your QC vial. A wrong value makes a good QC vial fail.

Did the blank fail, with code 5010 or 5011? Skip this step.

See Run QC test for where to find the value.

2

Use SBT consumables only

Get these ready, all at room temperature:

3

Run the QC test

Run the QC test with these consumables.

4

Read the result

  • QC test passes. A vial or liquid you used before caused the failure. Keep using SBT consumables. If you fill dilution vials yourself from SBT bulk liquid, replace the syringe filter at the start of each day, use SBT empty vials, and fill only the vials you use that day.

  • QC test fails. Contact support.

Was this helpful?