Recover a flow cell

Deep clean a flow cell that failed the QC test, then test it again.

Limescale and biofilm can build up in a flow cell over time, and make it fail the QC test. A deep clean can remove the build-up. Try the 15-minute deep clean first, and the 10-hour deep clean if that is not enough.

Do this when a new flow cell passed the QC test and the old one failed. See Check the flow cell.

Get things ready

You need these items:

Step by step

1

Install the flow cell that failed

Install the flow cell that failed the QC test.

2

Run a 15-minute deep clean

Run a 15-minute deep clean. Then run Clean with a fresh disinfection vial.

3

Run a QC test

Run a QC test with fresh vials. If it passes, the flow cell is ready to use again.

4

Run a 10-hour deep clean

QC test still fails? Run a 10-hour deep clean. Start it at the end of the day. Then run Clean with a fresh disinfection vial.

5

Run a QC test again

Run a QC test with fresh vials.

  • QC test passes. The flow cell is ready to use again.

  • QC test fails. Check the warranty on the flow cell page. If the flow cell is still under warranty, contact support. If not, use a new flow cell.

Always clean between a deep clean and a QC test