How-to
Cleaning liquid is not disinfectant
Last updated
Was this helpful?
You get error 1207 when the cleaning liquid conducts too much during a clean. The Disinfection vial holds a liquid that barely conducts. So a high conductivity means the liquid is not disinfectant, and your BactoBox® was not cleaned.
There are two common causes:
You loaded the wrong liquid. For example, you loaded dilution liquid instead of a disinfection vial.
You reused the same disinfection vial too many times. Each reuse adds a little conductivity. Use a disinfection vial at most 10 times.
Reuse adds conductivity over time. Use a disinfection vial at most 10 times. When in doubt, grab a fresh one. See Clean.
Did you try all of the above and nothing helped?
This is a hardware issue with your BactoBox® setup. Contact our support team to discuss the matter further. Make sure to mention it if you found:
A leak inside the flow cell compartment
Sign of corrosion or deterioration inside the flow cell department
Broken or loose inlet/outlet (where you connect the tubing kit to your BactoBox®)
Rattling or loose components
In general, hardware damage of any kind
If possible, include pictures of the problem. This helps us help you faster!