Download your data

BactoBox® appears as a USB-drive when turned on and connected to a Win7/Win10 PC with the included USB-cable. When clicking the USB-drive, a summary.csv file is available which can readily be imported to Excel. The “format as table” under the “Styles” pane works well for sorting data but file must be saved as e.g. .xlsx format to make changes persistent.

Field
Description

Program type

Specifies the nature of the performed program, e.g. a measurement program.

Start time

Format: YYYY-MM-DD hh:mm

Measurement ID

Unique measurement ID that was registered on the screen at the time of analysis.

Device ID

BactoBox® device ID

Software version

Software ID

Flow cell ID

Flow cell ID

Cells/mL

Cells/mL concentration

Total/mL

Total/mL concentration

Total/mL - Cells/mL

Other/mL concentraion

ECC factor

The empirical concentration correction factor used to adjust the intact cell concentration

Classification filename

Measurement program used. Typically default, unless special “gating” is applied

Classification limits

Phase and amplitude “gates” used for classifying the objects of interest.

Conductivity

Conductivity in µS/cm reported for the last 5 seconds of a measurement

Errors

Error codes and explanations when relevant

Last updated