signal-weakDetermine T0 concentration

Once the shake flask is inoculated, immediately collect a sample to obtain a measurement as close to T0 as possible.

1

Prepare the T0 sample for BactoBox® measurement

  1. Use a serological pipette to transfer a sample from the shake flask to a vial.

  2. Use a permanent marker to label the vial cap with the sampling time.

  3. Return the shake flask to the incubator shaker.

  4. Start a 30-minute countdown to remind you to collect the next sample.

  5. Vortex 30 seconds at max speed to disaggregate bacteria.

  6. Dilute 1:100 in diluent by adding 101 µL of sample to 10 mL of diluent. Vortex 10 sec., max speed.

  7. Transfer the tubing kit to the diluted sample.

circle-check

Put the flask back in the incubator shaker immediately after collecting the sample.

2

Measure the T0 sample in Access

  1. Click the Measure button in Access.

  2. Add metadata:

    1. Dilution: 100.

    2. Label: E_coli_growth.

  3. Once measurement is done, click the label E_coli_growth above the chart to visualize the growth curve data.

triangle-exclamation

Do you get the error conductivity too high?

3

Clean the device

  1. Transfer the tubing kit to the disinfection vial.

  2. Click Clean.

Summary

You now know the bacterial concentration at T0. Next we will collect a sample for every 30 min and keep repeating the 1:100 dilution scheme for lag and early exponential growth stage.