signal-bars-fairUse 1:100 dilutions initially

For every 30 minutes, collect a fresh sample and repeat the procedure from the 1:100 dilution.

circle-check

spray-can Do a Clean program after each measurement.

circle-check

hourglass Start a 30-minute countdown to remind you to collect the next sample.

1

Prepare the T0 sample for BactoBox® measurement

  1. Use a serological pipette to transfer a sample from the shake flask to a vial.

  2. Use a permanent marker to label the vial cap with the sampling time.

  3. Start a 30-minute countdown to remind you to collect the next sample.

  4. Vortex 30 seconds at max speed to disaggregate bacteria.

  5. Make a 1:100 dilution:

    1. Transfer 101 µL of your sample to 10 mL of diluent.

    2. Vortex 10 seconds at max speed.

  6. Transfer the tubing kit to the diluted sample.

2

Measure the diluted samples in Access

  1. Click the Measure button in Access.

  2. Add metadata:

    1. Dilution: 100.

    2. Label: E_coli_growth.

3

Clean the device

  1. Transfer the tubing kit to the disinfection vial.

  2. Click Clean.

Summary

The 1:100 dilution scheme works well for the measurements during lag and early-exponential growth stage. As the concentration increases, you'll need to dilute more. We will explain this in detail in the next section.

Last updated

Was this helpful?