For every 30 minutes, collect a fresh sample and repeat the procedure from the 1:100 dilution.
Do a Clean program after each measurement.
Start a 30-minute countdown to remind you to collect the next sample.
1
Prepare the T0 sample for BactoBox® measurement
Use a serological pipette to transfer a sample from the shake flask to a vial.
Use a permanent marker to label the vial cap with the sampling time.
Start a 30-minute countdown to remind you to collect the next sample.
Vortex 30 seconds at max speed to disaggregate bacteria.
Make a 1:100 dilution:
Transfer 101 µL of your sample to 10 mL of diluent.
Vortex 10 seconds at max speed.
Transfer the tubing kit to the diluted sample.
2
Measure the diluted samples in Access
Click the Measure button in Access.
Add metadata:
Dilution: 100.
Label: E_coli_growth.
3
Clean the device
Transfer the tubing kit to the disinfection vial.
Click Clean.
Summary
The 1:100 dilution scheme works well for the measurements during lag and early-exponential growth stage. As the concentration increases, you'll need to dilute more. We will explain this in detail in the next section.