Prepare your sample
Homogenize your sample as much as possible. Here are some options:
Agitate with magnetic stirrer
Dilute your sample
Prepare two dilution vials: Take two empty vials and add 10 mL of diluent to each.
1st dilution (total 10-2)
Dilute 1:100: Transfer 101 μL of your sample into the first dilution vial. Vortex (15 sec., Max RPM).
2nd dilution (total 10-4)
Dilute 1:100: Transfer 101 μL of from the first dilution vial into the second dilution vial. Vortex.
Measure your sample
Measure with BactoBox®. Start with the least concentrated sample and move towards higher concentrations.
First, clean your BactoBox® setup with a fresh disinfection vial.
Measure on the 10-4-diluted sample
If there result is above the limit of detection (above 30 000 cells/mL) then you are done! Otherwise, move on.
Measure on the 10-2-diluted sample
Is even the most-concentrated sample below the limit of detection? See Concentrate your sample for next steps.