Coarse samples
How to measure on coarse samples and with BactoBox®



Use BactoBox® to measure on stool samples, crude fermentation samples, skin swabs, heavily aggregated bacteria, pellicle-forming bacteria, etc. You must homogenize and filter these kind of samples first. Otherwise, it is literally shit in, shit out.
See What can I measure on? and Prepare your sample for details.
Overall approach
We separate bacteria from other particles in two steps:
Homogenize the sample
Filter the sample with a 5 or 10 µm nylon cell sieve
It is a fine-grained filter because BactoBox® does not count particles larger than 5 µm anyways (see Particle size).
Use live/dead fluorescence microscopy
Use a fluorescence microscope and the LIVE/DEAD™ BacLight™ staining kit to ensure that:
Bacteria are in single-cell form (no aggregates)
Bacteria are separate from other particles
Bacterial cell membrane is intact
Detailed protocol
Disaggregate bacteria and particles
Fill a flip-top tube with 9 mL buffered peptone water.
Transfer 1 mL or 1 g of sample to the flip-top tube.
Add 5 mm steel beads until the liquid level rises by 1 mL (typically 12 beads or 5 g).
Homogenize aggressively by bead beating (Max speed, Bead genie, 1–3 min).
In some cases, vortex with glass beads (3 mm, 3 g) is sufficient.
Use live/dead fluorescence microscopy
Use a fluorescence microscope and the LIVE/DEAD™ BacLight™ staining kit to ensure that:
Bacteria are in single-cell form (no aggregates)
Bacteria are separate from other particles
Bacterial cell membrane is intact
Remove large particles
Assemble a 3 mL syringe with a 10 µm cell sieve.
Remove the piston from the syringe and decant 3 mL homogenized sample into the syringe.
Filter approximately 1–2 mL sample to an empty vial.
Dilute your sample
Prepare two dilution vials: Take two empty vials and add 10 mL of diluent to each.
1st dilution (total 10-2)
Dilute 1:100: Transfer 101 μL of your sample into the first dilution vial. Vortex (15 sec., Max RPM).
2nd dilution (total 10-4)
Dilute 1:100: Transfer 101 μL of from the first dilution vial into the second dilution vial. Vortex.
Measure your sample
Measure with BactoBox®. Start with the least concentrated sample and move towards higher concentrations.
First, clean your BactoBox® setup with a fresh disinfection vial.
Measure on the 10-4-diluted sample
If there result is above the limit of detection (above 30 000 cells/mL) then you are done! Otherwise, move on.
Measure on the 10-2-diluted sample
Is even the most-concentrated sample below the limit of detection? See Concentrate your sample for next steps.
Multiply the cells/mL concentration with the dilution factor to calculate the cells/mL concentration in your sample stock.
Multiply the cells/mL concentration in your sample stock with the density (in g/mL) to calculate cells/g of your sample stock.
Last updated
Was this helpful?