Phase A: QC vial

To streamline things, BactoBox® both tunes and qualifies the setup simultaneously here in phase A

In practice, the QC test program implements most of phase A. However, there is a manual check against the certificate of analysis at the end. Be aware of that!

Quick clip

Get things ready

You need these items:

Clean the BactoBox® setup first

Step by step

1

Switch to the QC vial

Transfer tubing kit to the QC vial.

2

Start the QC test program

We do so with three button presses.

Enter the Programs menu

First, press and hold Measure until the Programs menu appears.

Press and hold Measure until the Program menu appears

Enter the QC test screen

Second, press Measure again briefly to select the QC test program.

Press Measure again briefly to select QC test program. The user interface shows screen like the above.

Start the QC test program

Third, press Measure again briefly to start the QC test program.

Press Measure again briefly to start the QC test program

After the final button press, BactoBox® starts the tune and qualify procedure.

3

Check the certificate of analysis

Wait for the QC test program to complete. Read the total/ml concentration.

See also Example A.

Qualification: Phase A passes

Qualification: Phase A fails

BactoBox® automatically checks cells/ml and gives an error if there are too many cells.

4

That is it!

During the QC test program, BactoBox® automatically tuned the flow parameters.

Discard the QC vial. It is for single use only.

Examples

The check the certificate of analysis step is a bit complex. Therefore, we provide a couple of examples below to help you along the way.

Example A

The QC test program results in 1.91e6 total/mL. The QC stock certificate of analysis has an accepted range of 1.62e6–2.19e6 total/mL. The qualification passes since the total/mL is within the accepted range.