Phase B: Dilution vial
In phase A (Phase A: QC vial) we measured the QC vial. Here in phase B, we measure the dilution vial directly. This qualifies that BactoBox® works both when there are particles and when there are not.
Quick clip ⚡

Get things ready
You need these items
1 × clean BactoBox® setup (see Clean the setup)
1 × dilution vial
If you came directly from phase A then you need to clean the BactoBox® setup first! Otherwise, there may be carry-over from the QC vial.
Clean the BactoBox® setup first
Make sure that you use a clean BactoBox® setup. See Clean the setup.
Step by step
Check the measurement result
Wait for the Measure program to complete. Read the cells/ml and total/ml concentrations.
Qualification: Phase B passes
The concentrations are below 30 000 cells/ml and below 100 000 total/ml.
Qualification: Phase B fails
If any error occurs, or, if the concentrations are outside the requirements listed above.
See Phase B: Dilution vial fails for a troubleshooting guide.