Phase B: Dilution vial

In phase A (Phase A: QC vial) we measured the QC vial. Here in phase B, we measure the dilution vial directly. This qualifies that BactoBox® works both when there are particles and when there are not.

Quick clip

Get things ready

You need these items

If you came directly from phase A then you need to clean the BactoBox® setup first! Otherwise, there may be carry-over from the QC vial.

Clean the BactoBox® setup first

Step by step

1

Switch to dilution vial

Transfer tubing kit to the dilution vial.

Transfer tubing kit to the dilution vial
2

Measure the dilution vial

Press Measure to start a measurement.

Press the Measure button
3

Check the measurement result

Wait for the Measure program to complete. Read the cells/ml and total/ml concentrations.

Qualification: Phase B passes

Qualification: Phase B fails

4

That is it!

If both phase A and phase B passed then your entire BactoBox® setup passes qualification.

Discard the dilution vial. It is for single use only.